Beam model development and clinical experience with RadCalc for treatment plan quality assurance in online adaptive workflow with an MR‐linac

Urszula Jelen, Zoë Moutrie, Jack D Aylward iD, Michael G Jameson

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70125 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Urszula Jelen und 3 weiteren Autor:innen. Beteiligte Einrichtung: GenesisCare, Alexandria, New South Wales, Australia. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Quality Assurance, Health Care, Phantoms, Imaging, Particle Accelerators. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The aim of this work was to report on the optimization, commissioning, and validation of a beam model using a commercial independent dose verification software RadCalc version 7.2 (Lifeline Software Inc, Tyler, TX, USA), along with 4 years of experience employing RadCalc for offline and online monitor unit (MU) verification on the Elekta Unity MR‐linac (MRL) for a range of clinical sites.

Methods

Calculation settings and model parameters, including the Clarkson integration settings and radiation/light field offset, have been systematically examined and optimized, and pitfalls in the use of density inhomogeneity corrections and in off‐axis calculations were investigated and addressed. The resulting model was commissioned by comparing RadCalc calculations to measurements for a variety of cases, selected following relevant recommendations, ranging from simple fields in a water tank to end‐to‐end point dose measurements in an anthropomorphic phantom.

Results

For simple geometries, the agreement was within 2%, and for complex geometries, within 5%. When validating against the Monaco (Elekta AB, Stockholm, Sweden) treatment planning system (TPS), for 39 clinical commissioning plans, the mean total point dose difference was −0.3 ± 0.8% (−2.0%–1.1%). Finally, when applied retrospectively to 4085 clinical plan calculations, the agreement with the TPS was 0.3 ± 1.1% (−4.8%–4.2%), with fail rates of 0.1% for total point dose (discrepancy > 4%) and 0.3% for individual fields (discrepancy > 10%).

Conclusions

Improved calculation agreement with the TPS and therefore increased confidence in the online QA, opened the way for an automated and physics‐light independent MU verification workflow within our MRL program.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedQuality Assurance, Health CarePhantoms, ImagingParticle AcceleratorsSoftwareNeoplasmsMagnetic Resonance ImagingHumansRadiotherapy DosageWorkflowRadiotherapy, Intensity-ModulatedAlgorithms

Stichwörter der Autor:innen

MR‐linacRadCalcadaptive radiotherapyindependent MU verificationpatient specific quality assurance

Autor:innen und Institutionen

  • GenesisCare, Alexandria, New South Wales, Australia.
  • South Western Sydney Cancer Services, New South Wales Health, Sydney, New South Wales, Australia.
  • University of New South Wales, Sydney, New South Wales, Australia.
  • Ingham Institute for Applied Medical Research, Liverpool, New South Wales, Australia.
  • Medical Physics, School of Applied Sciences, University of the West of England, Bristol, UK.
  • Medical Physics, University Hospitals Bristol and Weston NHS Foundation Trust, Bristol, UK.

ORCID-Profile: Jack D Aylward

Zitieren

APA

Jelen, U., Moutrie, Z., Aylward, J. D., & Jameson, M. G. (2025). Beam model development and clinical experience with RadCalc for treatment plan quality assurance in online adaptive workflow with an MR‐linac. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70125

Vancouver

Jelen U, Moutrie Z, Aylward JD, Jameson MG. Beam model development and clinical experience with RadCalc for treatment plan quality assurance in online adaptive workflow with an MR‐linac. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70125

BibTeX
@article{Jelen2025,
  author = {Jelen, Urszula and Moutrie, Zoë and Aylward, Jack D and Jameson, Michael G},
  title = {Beam model development and clinical experience with RadCalc for treatment plan quality assurance in online adaptive workflow with an MR‐linac},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70125}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche