EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Application of a model‐based water‐equivalent EPID image conversion algorithm for linac beam QA

Ivan Kutuzov iD, Boyd McCurdy

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70210 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Ivan Kutuzov und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, Canada. Hauptthemen laut MEDLINE: Algorithms, Particle Accelerators, Radiotherapy Planning, Computer-Assisted, Water. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The nonwater‐equivalent energy response of electronic portal imaging devices (EPIDs) is a major obstacle to using them for linear accelerator (linac) beam parameter verification. In this study, we propose an EPID‐based machine quality assurance (QA) application that uses a model‐based radiation transport algorithm to convert EPID‐measured images into water‐equivalent dose distributions that can be used to assess beam flatness and symmetry.

Methods

An in‐house developed, model‐based radiation transport algorithm was used to estimate the incident beam fluence from measured EPID images and convert it into either 3D dose distributions in a virtual water tank or 2D water‐equivalent dose distributions in a virtual ion chamber array. The conversion algorithm was validated using independent measurements in a scanning water tank and a reference ion chamber array under symmetric and also intentionally detuned (i.e., asymmetric) beam conditions.

Results

For symmetric fields, EPID‐reconstructed percentage depth dose distributions (PDDs) agreed with water tank measurements to within 1% beyond the first 10 mm of depth. Beam profile comparisons showed differences within 1% in low dose‐gradient regions. For all symmetric and intentionally asymmetric fields, beam flatness and symmetry derived from reconstructed images agreed with reference measurements to within 0.2% and 0.3%, respectively. The model demonstrated high sensitivity to the controlled beam asymmetries and steering distortions, with EPID‐reconstructed metrics closely matching reference water‐equivalent measurements and significantly outperforming metrics derived from raw EPID images.

Conclusions

The proposed model‐based algorithm enables accurate conversion of EPID images into water‐equivalent dose distributions, facilitating accurate determination of beam flatness and symmetry. This application addresses some limitations of the previously proposed EPID‐based linac QA techniques, which are limited to nonwater‐equivalent constancy checks, and supports the use of EPIDs as robust dosimetry tools for linac radiation beam parameter verification.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsParticle AcceleratorsRadiotherapy Planning, Computer-AssistedWaterQuality Assurance, Health CarePhantoms, ImagingImage Processing, Computer-AssistedRadiometryHumansRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

EPIDEPID dosimetrylinac QAmachine QAwater‐equivalent EPID

Substanzen

Water

Autor:innen und Institutionen

  • Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, Canada.
  • Medical Physics Department, CancerCare Manitoba, Winnipeg, Manitoba, Canada.
  • Department of Radiology, University of Manitoba, Winnipeg, Manitoba, Canada.

ORCID-Profile: Ivan Kutuzov

Zitieren

APA

Kutuzov, I., & McCurdy, B. (2025). Application of a model‐based water‐equivalent EPID image conversion algorithm for linac beam QA. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70210

Vancouver

Kutuzov I, McCurdy B. Application of a model‐based water‐equivalent EPID image conversion algorithm for linac beam QA. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70210

BibTeX
@article{Kutuzov2025,
  author = {Kutuzov, Ivan and McCurdy, Boyd},
  title = {Application of a model‐based water‐equivalent EPID image conversion algorithm for linac beam QA},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70210}
}
Kostenlos starten →