EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Validation of an in vivo transit dosimetry algorithm using Monte Carlo simulations and ionization chamber measurements

David Sánchez-Artuñedo, Savannah Pié-Padró, Marcelino Hermida-López, Maria Amor Duch-Guillén, Mercè Beltran-Vilagrasa

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14187 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von David Sánchez-Artuñedo und 4 weiteren Autor:innen. Beteiligte Einrichtung: Servei de Física i Protecció Radiològica, Hospital Universitari Vall d'Hebron, Barcelona, Spain. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 70 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Transit dosimetry is a safety tool based on the transit images acquired during treatment. Forward‐projection transit dosimetry software, as PerFRACTION, compares the transit images acquired with an expected image calculated from the DICOM plan, the CT, and the structure set. This work aims to validate PerFRACTION expected transit dose using PRIMO Monte Carlo simulations and ionization chamber measurements, and propose a methodology based on MPPG5a report.

Methods

The validation process was divided into three groups of tests according to MPPG5a: basic dose validation, IMRT dose validation, and heterogeneity correction validation. For the basic dose validation, the fields used were the nine fields needed to calibrate PerFRACTION and three jaws‐defined. For the IMRT dose validation, seven sweeping gaps fields, the MLC transmission and 29 IMRT fields from 10 breast treatment plans were measured. For the heterogeneity validation, the transit dose of these fields was studied using three phantoms: 10 , 30 , and a 3 cm cork slab placed between 10 cm of solid water. The PerFRACTION expected doses were compared with PRIMO Monte Carlo simulation results and ionization chamber measurements.

Results

Using the 10 cm solid water phantom, for the basic validation fields, the root mean square (RMS) of the difference between PerFRACTION and PRIMO simulations was 0.6%. In the IMRT fields, the RMS of the difference was 1.2%. When comparing respect ionization chamber measurements, the RMS of the difference was 1.0% both for the basic and the IMRT validation. The average passing rate with a γ(2%/2 mm, TH = 20%) criterion between PRIMO dose distribution and PerFRACTION expected dose was 96.0% ± 5.8%.

Conclusion

We validated PerFRACTION calculated transit dose with PRIMO Monte Carlo and ionization chamber measurements adapting the methodology of the MMPG5a report. The methodology presented can be applied to validate other forward‐projection transit dosimetry software.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedHumansRadiotherapy DosageMonte Carlo MethodParticle AcceleratorsAlgorithmsRadiometryPhantoms, ImagingWater

Stichwörter der Autor:innen

Monte CarloPRIMOPerFRACTIONtransit dosimetryvalidation

Substanzen

Water

Autor:innen und Institutionen

  • Servei de Física i Protecció Radiològica, Hospital Universitari Vall d'Hebron, Barcelona, Spain.
  • Institut de Tècniques Energètiques, Universitat Politècnica de Catalunya, Barcelona, Spain.

Zitieren

APA

Sánchez-Artuñedo, D., Pié-Padró, S., Hermida-López, M., Duch-Guillén, M. A., & Beltran-Vilagrasa, M. (2024). Validation of an in vivo transit dosimetry algorithm using Monte Carlo simulations and ionization chamber measurements. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14187

Vancouver

Sánchez-Artuñedo D, Pié-Padró S, Hermida-López M, Duch-Guillén MA, Beltran-Vilagrasa M. Validation of an in vivo transit dosimetry algorithm using Monte Carlo simulations and ionization chamber measurements. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14187

BibTeX
@article{SnchezArtuedo2024,
  author = {Sánchez-Artuñedo, David and Pié-Padró, Savannah and Hermida-López, Marcelino and Duch-Guillén, Maria Amor and Beltran-Vilagrasa, Mercè},
  title = {Validation of an in vivo transit dosimetry algorithm using Monte Carlo simulations and ionization chamber measurements},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14187}
}
Kostenlos starten →