Accounting for TLD response to CBCT protocols in external beam radiotherapy dose monitoring
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70103 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Andrew J White und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Thermoluminescent Dosimetry, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
LiF thermoluminescent dosimeters (TLDs) are commonly used in radiation therapy to verify the delivered dose. Examples include dose verification for complicated treatment setups or cardiovascular implantable electronic devices (CIEDs). TLDs may be present for both the pre‐treatment imaging kilovoltage (kV) beam and the megavoltage (MV) treatment beams. For low energy x‐ray beams, where the photoelectric effect dominates, TLDs respond differently than tissue or water. An overresponse of up to 40% has been previously reported for lower‐energy (kV) x‐rays when calibrated to higher‐energy (MV) beams. In this work, the response of TLDs to various cone beam computed tomography (CBCT) protocols with calibrations in clinical therapy beams (MV) is quantified.
Methods
Three Varian OBI (Head, Thorax, and Pelvis) and three Elekta XVI (Fast Head and Neck, Fast Chest, and Fast Pelvis) CBCT protocols were investigated. For each protocol, TLDs were positioned on a tissue equivalent phantom at various distances extending out from the center of the imaging field. The response was determined by calibrating TLDs to a 6 MV photon beam.
Results
The maximum in‐field TLD response ranged from 0.82 to 4.80 and 0.06 to 2.69 cGy for Varian OBI and Elekta XVI protocols, respectively. The out‐of‐field CBCT response dropped exponentially from the field edge. Calculated uncertainties were generally less than 3% ( k = 1), with exceptions along the edge of the CBCT field (6%) and at the most distal TLD positions (34%).
Conclusions
Using the measured TLD responses to CBCT protocols with an MV calibration, the therapeutic dose can be isolated. The therapeutic dose can then be compared to predictions from the treatment planning system (TPS), allowing for more accurate dose verification for complex treatment setups and patients with CIEDs. The CBCT response can change the reported therapeutic dose by up to 2.5%.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- Department of Radiation Oncology, Mercyhealth Javon Bea Hospital, Rockford, Illinois, USA.
- Department of Radiation Oncology, Mission Cancer Center, Asheville, North Carolina, USA.
ORCID-Profile: Andrew J White
Zitieren
APA
White, A. J., Hammer, C. G., Brenner, M. W., DeWerd, L. A., Stump, K. E., & Culberson, W. S. (2025). Accounting for TLD response to CBCT protocols in external beam radiotherapy dose monitoring. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70103
Vancouver
White AJ, Hammer CG, Brenner MW, DeWerd LA, Stump KE, Culberson WS. Accounting for TLD response to CBCT protocols in external beam radiotherapy dose monitoring. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70103
BibTeX
@article{White2025,
author = {White, Andrew J and Hammer, Cliff G and Brenner, Matthew W and DeWerd, Larry A and Stump, Kurt E and Culberson, Wesley S},
title = {Accounting for TLD response to CBCT protocols in external beam radiotherapy dose monitoring},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70103}
}
EUVIMED