EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Feasibility study of skin dosimetry with TLD sheets for measuring the effect of 3D printed bolus in radiotherapy

Yuya Miyasaka, Mayumi Ichikawa, Takagi Akira, Yoshifumi Yamazawa, Hongbo Chai, Hikaru Souda, Miyu Ishizawa iD, Hiraku Sato iD, Takeo Iwai

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70035 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Yuya Miyasaka und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Heavy Particle Medical Science, Yamagata University Graduate School of Medical Science, Yamagata, Japan. Hauptthemen laut MEDLINE: Printing, Three-Dimensional, Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging, Thermoluminescent Dosimetry. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The thermoluminescence dosimeter (TLD) sheet is a measurement device coated with manganese‐doped LiB 3 O 5 in sheet form. The sheet is 0.2‐mm thick and flexible. Hence, it can fit and be installed on irregular surfaces. The current study aimed to evaluate the feasibility of measuring patient surface doses during radiation therapy using TLD sheets. Further, the surface doses when using a three‐dimensional (3D) printed bolus were compared. First, calibration of the between TLD sheet measurements and the irradiation doses was performed. Second, dose calculations and TLD sheet measurements were compared to evaluate the error between treatment planning system (TPS) dose calculations and surface dose measurements. Finally, the TLD sheet was fitted to the head phantom to measure the surface dose at the uneven areas under two conditions (with a commercial [CM] bolus and a 3D‐printed bolus). Based on the error from the dose calculation algorithm and the difference between the TLD sheet thickness and the dose grid size or evaluation structure volume, the differences between the TLD sheet measurements and the treatment plan doses were up to 68.6% for the X‐ray collapsed cone convolution (CCC) method and up to 8.4% for the electron beam Monte Carlo (MC) method. Based on the results of the surface dosimetry of the head phantom, the application of surface dosimetry, particularly skin dosimetry, even on uneven body surfaces, can be feasible.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Printing, Three-DimensionalRadiotherapy Planning, Computer-AssistedPhantoms, ImagingThermoluminescent DosimetrySkinHumansFeasibility StudiesRadiotherapy DosageRadiotherapy, Intensity-ModulatedAlgorithms

Stichwörter der Autor:innen

3D printerTLD sheetbolusradiotherapysurface dose

Autor:innen und Institutionen

  • Department of Heavy Particle Medical Science, Yamagata University Graduate School of Medical Science, Yamagata, Japan.
  • Department of Radiology, Yamagata University Faculty of Medicine, Yamagata, Japan.
  • Division of Dentistry and Oral Surgery, Yamagata University Faculty of Medicine, Yamagata, Japan.
  • Department of Radiology, Yamagata University Hospital, Yamagata, Japan.

ORCID-Profile: Miyu Ishizawa, Hiraku Sato

Zitieren

APA

Miyasaka, Y., Ichikawa, M., Akira, T., Yamazawa, Y., Chai, H., Souda, H., Ishizawa, M., Sato, H., & Iwai, T. (2025). Feasibility study of skin dosimetry with TLD sheets for measuring the effect of 3D printed bolus in radiotherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70035

Vancouver

Miyasaka Y, Ichikawa M, Akira T, Yamazawa Y, Chai H, Souda H, et al. Feasibility study of skin dosimetry with TLD sheets for measuring the effect of 3D printed bolus in radiotherapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70035

BibTeX
@article{Miyasaka2025,
  author = {Miyasaka, Yuya and Ichikawa, Mayumi and Akira, Takagi and Yamazawa, Yoshifumi and Chai, Hongbo and Souda, Hikaru and Ishizawa, Miyu and Sato, Hiraku and Iwai, Takeo},
  title = {Feasibility study of skin dosimetry with TLD sheets for measuring the effect of 3D printed bolus in radiotherapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70035}
}
Kostenlos starten →