EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Risk analysis of the Unity 1.5T MR‐Linac adapt‐to‐shape workflow

Jiayi Liang iD, Eric Aliotta, Neelam Tyagi, Paola Godoy Scripes, Nicolas Côté, Ergys Subashi, Qijie Huang, Lian Sun, Ching-Yun Chan, Angela Ng, Theresa Wunner, Victoria Brennan und 2 weitere

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70095 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Jiayi Liang und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, USA. Hauptthemen laut MEDLINE: Workflow, Software, Magnetic Resonance Imaging, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Purpose The adapt‐to‐shape (ATS) workflow on the Unity MR‐Linac (Elekta AB, Stockholm, Sweden) allows for full replanning including recontouring and reoptimization 5 . Additional complexity to this workflow is added when the adaptation involves the use of MIM Maestro (MIM Software, Cleveland, OH) software in conjunction with Monaco (Elekta AB, Stockholm, Sweden). Given the interplay of various systems and the inherent complexity of the ATS workflow, a risk analysis would be instructive.

Method

Failure modes and effects analysis (FMEA) following Task Group 100 13 was completed to evaluate the ATS workflow. A multi‐disciplinary team was formed for this analysis. The team created a process map detailing the steps involved in ATS treating both the standard Monaco workflow and a workflow with the use of MIM software in parallel. From this, failure modes were identified, scored using three categories (likelihood of occurrence, severity, and detectability which multiplied create a risk priority number), and then mitigations for the top 20 th percentile of failure modes were found.

Results

Risk analysis found 264 failure modes in the ATS workflow. Of those, 82 were high‐ranking failure modes that ranked in the top 20 th percentile for risk priority number and severity scores. Although high‐ranking failure modes were identified in each step in the process, 62 of them were found in the contouring and planning steps, highlighting key differences from adapt‐to‐position (ATP), where the importance of these steps are minimized. Mitigations are suggested for all high‐ranking failure modes.

Conclusion

The flexibility of the ATS workflow, which enables reoptimization of the treatment plan, also introduces potential critical points where errors can occur. There are more opportunities for error in ATS that can create unintentionally negative dosimetric impact. FMEA can help mitigate these risks by identifying and addressing potential failure points in the ATS process.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

WorkflowSoftwareMagnetic Resonance ImagingRadiotherapy Planning, Computer-AssistedRadiotherapy, Intensity-ModulatedImage Processing, Computer-AssistedPhantoms, ImagingNeoplasmsHumansRisk AssessmentRadiotherapy DosageHealthcare Failure Mode and Effect Analysis

Stichwörter der Autor:innen

MR‐Linacadaptive treatmentadapt‐to‐shapeprocess mapradiation oncologyrisk analysis

Autor:innen und Institutionen

  • Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, USA.
  • Department of Radiation Oncology, Memorial Sloan Kettering Cancer Center, New York, USA.

ORCID-Profile: Jiayi Liang

Förderung

  • NCI NIH HHS

Zitieren

APA

Liang, J., Aliotta, E., Tyagi, N., Godoy Scripes, P., Côté, N., Subashi, E., Huang, Q., Sun, L., Chan, C. Y., Ng, A., Wunner, T., Brennan, V., Zakeri, K., & Mechalakos, J. (2025). Risk analysis of the Unity 1.5T MR‐Linac adapt‐to‐shape workflow. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70095

Vancouver

Liang J, Aliotta E, Tyagi N, Godoy Scripes P, Côté N, Subashi E, et al. Risk analysis of the Unity 1.5T MR‐Linac adapt‐to‐shape workflow. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70095

BibTeX
@article{Liang2025,
  author = {Liang, Jiayi and Aliotta, Eric and Tyagi, Neelam and Godoy Scripes, Paola and Côté, Nicolas and Subashi, Ergys and Huang, Qijie and Sun, Lian and Chan, Ching-Yun and Ng, Angela and Wunner, Theresa and Brennan, Victoria and Zakeri, Kaveh and Mechalakos, James},
  title = {Risk analysis of the Unity 1.5T MR‐Linac adapt‐to‐shape workflow},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70095}
}
Kostenlos starten →