Fachartikel in der Zeitschrift Technology in Cancer Research & Treatment, 2026. Verfasst von Lingling Yan und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sci…. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Magnetic Resonance Imaging, Radiotherapy, Image-Guided, Neoplasms. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: CAMS Innovation Fund for Medical Sciences, National Natural Science Foundation, National High Level Hospital Clinical Research Funding and Cooperation Fund of CHCAMS Beijing & Langfang & SZCH.
Abstract (Originalsprache)
Introduction
Owing to the limitation in the field size of the magnetic resonance (MR)-Linac, currently, tumors with a length of >20 cm cannot be treated. Thus, the present study aimed to develop an expanded magnetic resonance imaging-guided adaptive radiotherapy (MRIgART) workflow for long, continuous planning target volumes (PTVs).
Methods
The PTVs were divided into two sub_target volumes (PTV_sub1 and PTV_sub2). We established two isocenters and defined a field overlap region. By adjusting the MR scan range, devising the online and offline adaptive procedures, synchronizing the online adaptive processes, and constructing a pretreatment dose evaluation, a new MRIgART workflow for long PTVs was established. The new workflow was validated using an in-house-made MR phantom. Additionally, the ArcherQA Monte Carlo-based method, ArcCHECK phantom, and ionization chamber measurement method were used for dose verification.
Results
Two clinical scenarios were established: (1) both PTV_sub1 and PTV_sub2 followed the adapt-to-position (ATP) workflow, and (2) PTV_sub1 followed the adapt-to-shape (ATS) workflow, whereas PTV_sub2 followed the ATP workflow. The feasibility of the proposed MRIgART workflow for long, continuous PTVs was demonstrated through three independent rounds of testing and validation for each scenario. When field overlaps were utilized, the PTV length that can be treated is 40 cm minus the length of field overlap region. The average gamma pass rates for the PTV_sub1 and PTV_sub2 adaptive plans were 95.74% and 98.63%, respectively (ArcherQA vs TPS). For the field overlap region, the average gamma pass rate was 95.50% (ArcCHECK vs TPS). The difference between the ionization chamber measurements and calculated results was smaller than 2%.
Conclusion
This study demonstrated the feasibility, safety, and accuracy of the MRIgART workflow for long PTVs. This workflow provides an effective solution for expanding the application of MRIgART to patients with long, continuous PTVs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
ORCID-Profile: Jianrong Dai
Förderung
- CAMS Innovation Fund for Medical Sciences
- National Natural Science Foundation
- National High Level Hospital Clinical Research Funding and Cooperation Fund of CHCAMS Beijing & Langfang & SZCH
- Cancer Research Program of National Cancer Center
- Special Research Fund for Central Universities, Peking Union Medical College
Zitieren
APA
Yan, L., Wang, N., Zhang, K., Nie, W., Qin, S., Li, X., Chen, D., Fu, Q., Dai, J., & Men, K. (2026). Development and Validation of a Magnetic Resonance Imaging-Guided Adaptive Radiotherapy Workflow for Long, Continuous Planning Target Volumes. Technology in Cancer Research & Treatment. https://doi.org/10.1177/15330338251408324
Vancouver
Yan L, Wang N, Zhang K, Nie W, Qin S, Li X, et al. Development and Validation of a Magnetic Resonance Imaging-Guided Adaptive Radiotherapy Workflow for Long, Continuous Planning Target Volumes. Technology in Cancer Research & Treatment. 2026. doi:10.1177/15330338251408324
BibTeX
@article{Yan2026,
author = {Yan, Lingling and Wang, NingYu and Zhang, Ke and Nie, Wensheng and Qin, Shirui and Li, Xiufen and Chen, Deqi and Fu, Qi and Dai, Jianrong and Men, Kuo},
title = {Development and Validation of a Magnetic Resonance Imaging-Guided Adaptive Radiotherapy Workflow for Long, Continuous Planning Target Volumes},
journal = {Technology in Cancer Research & Treatment},
year = {2026},
doi = {10.1177/15330338251408324}
}
EUVIMED