A method for measuring spatial resolution based on clinical chest CT sequence images
Journal of Applied Clinical Medical Physics · 2025-05 · 26 (5)
DOI: 10.1002/acm2.70078 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 18.03.2025. Verfasst von Ying Liu und 3 weiteren Autor:innen. Beteiligte Einrichtung: School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, China. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Algorithms, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China, Shanghai Engineering Research Center of Interventional Medical Device.
Abstract (Originalsprache)
Purpose
This study aimed to develop and validate a method for characterizing the spatial resolution of clinical chest computed tomography (CT) sequence images.
Methods
An algorithm for characterizing spatial resolution based on clinical chest CT sequence images was developed in Matlab (2021b). The algorithm was validated using CT sequence images from a custom‐made chest automatic tube current modulation (ATCM) phantom and clinically reconstructed chest CT sequence images. A region of interest (ROI) was automatically established at the edges of CT image subject to calculate the edge spread function (ESF). The ESF curves from consecutive CT images within the same sequence were fitted into a curve, and the line spread function (LSF) was derived through differentiation. A Fourier transformation of the LSF curve was conducted to obtain the modulation transfer function (MTF). The method's effectiveness was verified by comparing the 50% MTF and 10% MTF values with those calculated using IndoQCT (22a) software. The method was also applied to clinical CT images to calculate MTF values for various reconstructions, confirming its sensitivity by determining spatial resolution of clinically reconstructed images.
Results
Validation experiments based on the phantom CT sequence images demonstrated that the MTF values calculated using the proposed method had an average difference of within ± 5% compared to the results obtained with IndoQCT. Validation experiments with clinical CT sequence images indicated that the method effectively reflects differences and variations in spatial resolution of images under different reconstruction kernels, with the MTF values for B10f‐B50f and D10f‐D50f exhibiting a consistent increase.
Conclusion
A method for measuring spatial resolution using clinical chest CT sequence images was developed. This method provides a direct means of spatial resolution characterization for clinical CT datasets and a more accurate representation of CT imaging quality, effectively reflects variations across different reconstruction convolution kernels, demonstrating its sensitivity.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Health Science and Engineering, University of Shanghai for Science and Technology, Shanghai, China.
- Institute of Radiation Medicine, Fudan University, Shanghai, China.
Förderung
- National Natural Science Foundation of China
- Shanghai Engineering Research Center of Interventional Medical Device
Zitieren
APA
Liu, Y., Shen, J., Zhang, H., & Liu, H. (2025). A method for measuring spatial resolution based on clinical chest CT sequence images. Journal of Applied Clinical Medical Physics, 26(5). https://doi.org/10.1002/acm2.70078
Vancouver
Liu Y, Shen J, Zhang H, Liu H. A method for measuring spatial resolution based on clinical chest CT sequence images. Journal of Applied Clinical Medical Physics. 2025;26(5). doi:10.1002/acm2.70078
BibTeX
@article{Liu2025,
author = {Liu, Ying and Shen, Jingying and Zhang, Haowei and Liu, Haikuan},
title = {A method for measuring spatial resolution based on clinical chest CT sequence images},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
volume = {26},
number = {5},
doi = {10.1002/acm2.70078}
}