Digital phantom versus patient‐specific radiation dosimetry in adult routine thorax CT examinations
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14389 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Antonios E Papadakis und 5 weiteren Autor:innen. Beteiligte Einrichtung: University Hospital of Heraklion, Medical Physics Department, Stavrakia, Heraklion, Crete, Greece. Hauptthemen laut MEDLINE: Phantoms, Imaging, Tomography, X-Ray Computed, Monte Carlo Method, Radiation Dosage. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Euratom Reasearch and Training Program, SINFONIA.
Abstract (Originalsprache)
Purpose
The aim of this study was to compare the organ doses assessed through a digital phantom‐based and a patient specific‐based dosimetric tool in adult routine thorax computed tomography (CT) examinations with reference to physical dose measurements performed in anthropomorphic phantoms.
Methods
Two Monte Carlo based dose calculation tools were used to assess organ doses in routine adult thorax CT examinations. These were a digital phantom‐based dosimetry tool (NCICT, National Cancer Institute, USA) and a patient‐specific individualized dosimetry tool (ImpactMC, CT Imaging GmbH, Germany). Digital phantoms and patients were classified in four groups according to their water equivalent diameter ( D w ). Normalized to volume computed tomography dose index (CTDI vol ), organ dose was assessed for lungs, esophagus, heart, breast, active bone marrow, and skin. Organ doses were compared to measurements performed using thermoluminescent detectors (TLDs) in two physical anthropomorphic phantoms that simulate the average adult individual as a male (Alderson Research Labs, USA) and as a female (ATOM Phantoms, USA).
Results
The average percent difference of NCICT to TLD and ImpactMC to TLD dose measurements across all organs in both sexes was 13% and 6%, respectively. The average ± 1 standard deviation in dose values across all organs with NCICT, ImpactMC, and TLDs was ± 0.06 (mGy/mGy), ± 0.19 (mGy/mGy), and ± 0.13 (mGy/mGy), respectively. Organ doses decreased with increasing D w in both NCICT and ImpactMC.
Conclusion
Organ doses estimated with ImpactMC were in closer agreement to TLDs compared to NCICT. This may be attributed to the inherent property of ImpactMC methodology to generate phantoms that resemble the realistic anatomy of the examined patient as opposed to NCICT methodology that incorporates an anatomical discrepancy between phantoms and patients.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- University Hospital of Heraklion, Medical Physics Department, Stavrakia, Heraklion, Crete, Greece.
- Division of Nuclear Medicine and Molecular Imaging, Geneva University Hospital, Geneva, Switzerland.
- Department of Nuclear Medicine and Molecular Imaging, University of Groningen, University Medical Center Groningen, Groningen, Netherlands.
- Department of Nuclear Medicine, University of Southern Denmark, Odense, Denmark.
- University Research and Innovation Center, Obuda University, Budapest, Hungary.
- University of Crete, Medical School, Medical Physics Department, Stavrakia, Heraklion, Crete, Greece.
ORCID-Profile: Habib Zaidi
Förderung
- Euratom Reasearch and Training Program
- SINFONIA
Zitieren
APA
Papadakis, A. E., Giannakaki, V., Stratakis, J., Myronakis, M., Zaidi, H., & Damilakis, J. (2024). Digital phantom versus patient‐specific radiation dosimetry in adult routine thorax CT examinations. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14389
Vancouver
Papadakis AE, Giannakaki V, Stratakis J, Myronakis M, Zaidi H, Damilakis J. Digital phantom versus patient‐specific radiation dosimetry in adult routine thorax CT examinations. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14389
BibTeX
@article{Papadakis2024,
author = {Papadakis, Antonios E and Giannakaki, Vassiliki and Stratakis, John and Myronakis, Marios and Zaidi, Habib and Damilakis, John},
title = {Digital phantom versus patient‐specific radiation dosimetry in adult routine thorax CT examinations},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14389}
}
EUVIMED