Multi‐institutional validation of hypersight CBCT‐based dose calculation on O‐ring linacs

Chih-Yuan Lin, Yi-Ling Chen, Chia-Chi Chang, Yin-Hsun Hu, Chia-Peng Pan, Fang-Hui Liu, Hsiang-Ping Chao, Yang-Wei Hsieh iD, Yu-Wei Lin iD, Chi-Yuan Yeh iD, Tzu-Yuan Chao, Shih-Ming Hsu

Journal of Applied Clinical Medical Physics · 2026-02 · 27 (2)

DOI: 10.1002/acm2.70512 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 16.02.2026. Verfasst von Chih-Yuan Lin und 11 weiteren Autor:innen. Beteiligte Einrichtung: Varian Medical Systems, A Siemens Healthineers Company, Taipei, Taiwan. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Particle Accelerators. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Conventional cone‐beam computed tomography (CBCT) systems are limited by suboptimal image quality, inaccurate Hounsfield unit (HU) calibration, and reduced reliability for dose calculation. HyperSight CBCT on the Halcyon platform offers improved HU accuracy, expanded field‐of‐view (FOV), and enhanced image quality.

Purpose

This study aimed to assess the dosimetric accuracy of treatment planning using HyperSight CBCT through phantom‐based dose verification.

Methods

This study included three institutions equipped with the HyperSight imaging system on the Halcyon platform, with all procedures performed after acceptance testing and calibration. Each institution generated HU‐to‐density calibration curves using computed tomography (CT) scanners and standardized phantoms, and corresponding CBCT for planning (CBCTp) scans were also acquired. Additional CBCTp scans were acquired using a consistent phantom model (062 M) across the three institutions. Reference treatment plans were created on CT images and transferred to CBCTp and CBCT datasets for dose recalculation using identical parameters. Dosimetric assessment included gamma analysis and comparisons of DVH‐based dosimetric metrics for relevant regions of interest (ROIs). End‐to‐end testing with an anthropomorphic phantom was performed using ion chamber measurements and film dosimetry at brain, bone, and thorax locations.

Results

HU‐to‐density curves showed consistent behavior across institutions, with larger variability only at higher densities. CBCTp calibrations agreed well with vendor references. DVH‐based dosimetric metrics showed differences generally within 1% for both CBCTp and CBCT when compared with CT. Across institutions, gamma analysis of both CBCTp and CBCT yielded high passing rates (≥ 98.5% at 3%/2 mm). End‐to‐end testing with film dosimetry showed that CBCT‐based plans agreed with measured doses within ± 4%, while CT‐based plans were within ± 3%. Ion chamber measurements showed all dose differences within ± 2.3%, with both CBCTp and CBCT within ± 1.0% of CT.

Conclusions

HyperSight CBCT provides accurate dose calculations when properly calibrated. Phantom‐based validation demonstrated sub‐2% deviations and strong agreement with CT, supporting its clinical use in adaptive radiotherapy.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyPhantoms, ImagingRadiotherapy Planning, Computer-AssistedParticle AcceleratorsImage Processing, Computer-AssistedHumansRadiotherapy DosageRadiotherapy, Intensity-ModulatedCalibrationOrgans at Risk

Stichwörter der Autor:innen

HU‐to‐density calibration curveO‐Ring Linacdose calculationdosimetric validationhypersight CBCTmulti‐institutional study

Autor:innen und Institutionen

  • Varian Medical Systems, A Siemens Healthineers Company, Taipei, Taiwan.
  • Department of Biomedical Imaging and Radiological Sciences, National Yang Ming Chiao Tung University, Taipei, Taiwan.
  • Medical Physics and Radiation Measurements Laboratory, National Yang Ming Chiao Tung University, Taipei, Taiwan.
  • Department of Radiation Oncology, Kaohsiung Veterans General Hospital, Kaohsiung, Taiwan.
  • Department of Radiation Oncology, Tungs' Taichung MetroHarbor Hospital, Taichung, Taiwan.
  • Department of Radiation Oncolongy, Kaohsiung Show Chwan Memorial Hospital, Kaohsiung, Taiwan.

ORCID-Profile: Yang-Wei Hsieh, Yu-Wei Lin, Chi-Yuan Yeh

Zitieren

APA

Lin, C. Y., Chen, Y. L., Chang, C. C., Hu, Y. H., Pan, C. P., Liu, F. H., Chao, H. P., Hsieh, Y. W., Lin, Y. W., Yeh, C. Y., Chao, T. Y., & Hsu, S. M. (2026). Multi‐institutional validation of hypersight CBCT‐based dose calculation on O‐ring linacs. Journal of Applied Clinical Medical Physics, 27(2). https://doi.org/10.1002/acm2.70512

Vancouver

Lin CY, Chen YL, Chang CC, Hu YH, Pan CP, Liu FH, et al. Multi‐institutional validation of hypersight CBCT‐based dose calculation on O‐ring linacs. Journal of Applied Clinical Medical Physics. 2026;27(2). doi:10.1002/acm2.70512

BibTeX
@article{Lin2026,
  author = {Lin, Chih-Yuan and Chen, Yi-Ling and Chang, Chia-Chi and Hu, Yin-Hsun and Pan, Chia-Peng and Liu, Fang-Hui and Chao, Hsiang-Ping and Hsieh, Yang-Wei and Lin, Yu-Wei and Yeh, Chi-Yuan and Chao, Tzu-Yuan and Hsu, Shih-Ming},
  title = {Multi‐institutional validation of hypersight CBCT‐based dose calculation on O‐ring linacs},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {2},
  doi = {10.1002/acm2.70512}
}
Kostenlos starten →