Multi‐institutional validation of hypersight CBCT‐based dose calculation on O‐ring linacs
Journal of Applied Clinical Medical Physics · 2026-02 · 27 (2)
DOI: 10.1002/acm2.70512 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 16.02.2026. Verfasst von Chih-Yuan Lin und 11 weiteren Autor:innen. Beteiligte Einrichtung: Varian Medical Systems, A Siemens Healthineers Company, Taipei, Taiwan. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Particle Accelerators. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Conventional cone‐beam computed tomography (CBCT) systems are limited by suboptimal image quality, inaccurate Hounsfield unit (HU) calibration, and reduced reliability for dose calculation. HyperSight CBCT on the Halcyon platform offers improved HU accuracy, expanded field‐of‐view (FOV), and enhanced image quality.
Purpose
This study aimed to assess the dosimetric accuracy of treatment planning using HyperSight CBCT through phantom‐based dose verification.
Methods
This study included three institutions equipped with the HyperSight imaging system on the Halcyon platform, with all procedures performed after acceptance testing and calibration. Each institution generated HU‐to‐density calibration curves using computed tomography (CT) scanners and standardized phantoms, and corresponding CBCT for planning (CBCTp) scans were also acquired. Additional CBCTp scans were acquired using a consistent phantom model (062 M) across the three institutions. Reference treatment plans were created on CT images and transferred to CBCTp and CBCT datasets for dose recalculation using identical parameters. Dosimetric assessment included gamma analysis and comparisons of DVH‐based dosimetric metrics for relevant regions of interest (ROIs). End‐to‐end testing with an anthropomorphic phantom was performed using ion chamber measurements and film dosimetry at brain, bone, and thorax locations.
Results
HU‐to‐density curves showed consistent behavior across institutions, with larger variability only at higher densities. CBCTp calibrations agreed well with vendor references. DVH‐based dosimetric metrics showed differences generally within 1% for both CBCTp and CBCT when compared with CT. Across institutions, gamma analysis of both CBCTp and CBCT yielded high passing rates (≥ 98.5% at 3%/2 mm). End‐to‐end testing with film dosimetry showed that CBCT‐based plans agreed with measured doses within ± 4%, while CT‐based plans were within ± 3%. Ion chamber measurements showed all dose differences within ± 2.3%, with both CBCTp and CBCT within ± 1.0% of CT.
Conclusions
HyperSight CBCT provides accurate dose calculations when properly calibrated. Phantom‐based validation demonstrated sub‐2% deviations and strong agreement with CT, supporting its clinical use in adaptive radiotherapy.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Varian Medical Systems, A Siemens Healthineers Company, Taipei, Taiwan.
- Department of Biomedical Imaging and Radiological Sciences, National Yang Ming Chiao Tung University, Taipei, Taiwan.
- Medical Physics and Radiation Measurements Laboratory, National Yang Ming Chiao Tung University, Taipei, Taiwan.
- Department of Radiation Oncology, Kaohsiung Veterans General Hospital, Kaohsiung, Taiwan.
- Department of Radiation Oncology, Tungs' Taichung MetroHarbor Hospital, Taichung, Taiwan.
- Department of Radiation Oncolongy, Kaohsiung Show Chwan Memorial Hospital, Kaohsiung, Taiwan.
ORCID-Profile: Yang-Wei Hsieh, Yu-Wei Lin, Chi-Yuan Yeh
Zitieren
APA
Lin, C. Y., Chen, Y. L., Chang, C. C., Hu, Y. H., Pan, C. P., Liu, F. H., Chao, H. P., Hsieh, Y. W., Lin, Y. W., Yeh, C. Y., Chao, T. Y., & Hsu, S. M. (2026). Multi‐institutional validation of hypersight CBCT‐based dose calculation on O‐ring linacs. Journal of Applied Clinical Medical Physics, 27(2). https://doi.org/10.1002/acm2.70512
Vancouver
Lin CY, Chen YL, Chang CC, Hu YH, Pan CP, Liu FH, et al. Multi‐institutional validation of hypersight CBCT‐based dose calculation on O‐ring linacs. Journal of Applied Clinical Medical Physics. 2026;27(2). doi:10.1002/acm2.70512
BibTeX
@article{Lin2026,
author = {Lin, Chih-Yuan and Chen, Yi-Ling and Chang, Chia-Chi and Hu, Yin-Hsun and Pan, Chia-Peng and Liu, Fang-Hui and Chao, Hsiang-Ping and Hsieh, Yang-Wei and Lin, Yu-Wei and Yeh, Chi-Yuan and Chao, Tzu-Yuan and Hsu, Shih-Ming},
title = {Multi‐institutional validation of hypersight CBCT‐based dose calculation on O‐ring linacs},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {2},
doi = {10.1002/acm2.70512}
}
EUVIMED