Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Chul Hang Kim und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Gyeongsang National University School of Medicine, and Gyeongsang National University Changwon Hospital, Changwon, …. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Radiotherapy Planning, Computer-Assisted, Prostatic Neoplasms, Quality Assurance, Health Care. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study aimed to compare the sensitivity of log file‐based quality assurance (LF‐QA) and measurement‐based quality assurance (MB‐QA) for detecting multileaf collimator (MLC) positional errors and to evaluate the dosimetric impacts of MLC mechanical drive train defects.

Methods

Mechanical degradation of the MLC was simulated on a TrueBeam STx system by inducing three defect types: T‐nut surface wear (0.5–1.2 mm), drive screw thread wear, and motor degradation. MLC positioning accuracy was assessed using a rotational Picket Fence (PF) test, and the dosimetric impacts were evaluated on clinical breast intensity‑modulated radiation therapy (IMRT) and prostate volumetric‑modulated arc therapy (VMAT) plans. LF‐QA and MB‐QA were performed concurrently under identical delivery conditions. Gamma passing rates (GPRs) and dose‐volume histogram (DVH) analyses were compared between baseline and defective deliveries.

Results

LF‐QA detected positional deviations between baseline and defective conditions (

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedProstatic NeoplasmsQuality Assurance, Health CareBreast NeoplasmsPhantoms, ImagingParticle AcceleratorsHumansRadiotherapy DosageMaleOrgans at RiskFemaleRadiometry

Stichwörter der Autor:innen

MLC positional accuracylog file‐based QAmeasurement‐based QAmechanical wear

Autor:innen und Institutionen

  • Department of Radiation Oncology, Gyeongsang National University School of Medicine, and Gyeongsang National University Changwon Hospital, Changwon, Republic of Korea.
  • Department of Biomedical Engineering, Pusan National University, Pusan, Republic of Korea.
  • Institute of Health Science, Gyeongsang National University, Jinju, Republic of Korea.

ORCID-Profile: Hoon Sik Choi

Zitieren

APA

Kim, C. H., Kang, K. M., & Choi, H. S. (2025). Comparison of log file‐based and measurement‐based QA for detecting MLC positional errors and evaluating dosimetric impacts of MLC defects. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70389

Vancouver

Kim CH, Kang KM, Choi HS. Comparison of log file‐based and measurement‐based QA for detecting MLC positional errors and evaluating dosimetric impacts of MLC defects. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70389

BibTeX
@article{Kim2025,
  author = {Kim, Chul Hang and Kang, Ki Mun and Choi, Hoon Sik},
  title = {Comparison of log file‐based and measurement‐based QA for detecting MLC positional errors and evaluating dosimetric impacts of MLC defects},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70389}
}
Kostenlos starten →