EUVIMEDObjevte EUVIMED
Open access · CC BY 4.0

Commissioning of a commercial treatment planning system for scanned carbon‐ion radiotherapy

Wei Sun, Weiwei Wang iD, Zhijie Huang, Jingfang Zhao

Journal of Applied Clinical Medical Physics · 2025-03 · 26 (3)

DOI: 10.1002/acm2.14580 ↗

At a glance

Journal article in Journal of Applied Clinical Medical Physics, published online on 2024-11-29. Written by Wei Sun and 3 co-authors. Institution: Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Shanghai Key Laboratory of Radiation Oncology, Shanghai Engineering Research Cen…. Major MEDLINE topics: Radiotherapy Planning, Computer-Assisted, Phantoms, Imaging, Heavy Ion Radiotherapy, Algorithms. Cited 4 times so far (Crossref); 40 references. The full text is freely available (open access, CC BY 4.0). Funded by: Science and Technology Development Fund of Shanghai Pudong New Area, Natural Science Foundation of Shanghai, Natural Science Foundation of Shanghai Municipality.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

Purpose

To commission the RayStation (RS) TPS (treatment planning system) for scanned CIRT (carbon‐ion radiotherapy) utilizing pencil beam algorithms (PBv4.2).

Methods

The beam model commissioning entailed employing 1D single beams and 2D monoenergetic fields to validate spot profiles with films, assess beam range using Peakfinder measurements, and evaluate fragment spectra through dose‐averaged linear energy transfer (LETd) calculations. 3D dose distributions were verified in homogeneous phantoms for both absorbed and relative biological effectiveness (RBE)‐weighted doses, and further assessed in double wedge and anthropomorphic phantoms for absorbed dose only. Finally, RBE‐weighted dose verification and patient‐specific quality assurance were conducted using 58 beams from 20 clinically treated patient plans.

Results

The results demonstrated good agreement in absolute dose distribution between TPS calculations and measurements, with mean dose discrepancies within 3%. However, deviations were slightly higher (> 1%) for the cases involving the range shifter (RaShi) compared to those without the RaShi (< 1%). Beam range, depth dose distribution, and lateral profiles of spread‐out Bragg peaks (SOBPs) closely matched between RS TPS calculations and measurements. Some discrepancies (less than 0.5 mm) were observed at field edges and in penumbra regions due to limitations in simulating asymmetrical spots, but within clinical tolerance. After model tuning, RBE‐weighted dose calculations in RS TPS were in agreement with those from the clinically used TPS, except for variations exceeding 3% observed at energies exceeding 408.07 MeV/u, primarily attributed to fragment spectra differences.

Conclusion

Overall, this study validated the RS TPS for calculating absorbed doses against measurements and RBE‐weighted doses against a clinically used TPS. The results suggested that the RS TPS could be utilized for CIRT treatment planning, except for energies exceeding 408.07 MeV/u.

Subjects

MeSH terms (MEDLINE)

Radiotherapy Planning, Computer-AssistedPhantoms, ImagingHeavy Ion RadiotherapyAlgorithmsNeoplasmsQuality Assurance, Health CareHumansRadiotherapy DosageRelative Biological EffectivenessRadiotherapy, Intensity-ModulatedLinear Energy TransferOrgans at Risk

Author keywords

TPS commissioningcarbon ionspencil beam scanning

Authors and institutions

  • Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Shanghai Key Laboratory of Radiation Oncology, Shanghai Engineering Research Center of Proton and Heavy Ion Radiation Therapy, Shanghai, China.
  • Department of Medical Physics, Shanghai Proton and Heavy Ion Center, Fudan University Cancer Hospital, Shanghai, China.
  • Institute of Modern Physics, Applied Ion Beam Physics Laboratory, Fudan University, Shanghai, China.

ORCID profiles: Weiwei Wang

Funding

  • Science and Technology Development Fund of Shanghai Pudong New Area
  • Natural Science Foundation of Shanghai
  • Natural Science Foundation of Shanghai Municipality

Cite

APA

Sun, W., Wang, W., Huang, Z., & Zhao, J. (2025). Commissioning of a commercial treatment planning system for scanned carbon‐ion radiotherapy. Journal of Applied Clinical Medical Physics, 26(3). https://doi.org/10.1002/acm2.14580

Vancouver

Sun W, Wang W, Huang Z, Zhao J. Commissioning of a commercial treatment planning system for scanned carbon‐ion radiotherapy. Journal of Applied Clinical Medical Physics. 2025;26(3). doi:10.1002/acm2.14580

BibTeX
@article{Sun2025,
  author = {Sun, Wei and Wang, Weiwei and Huang, Zhijie and Zhao, Jingfang},
  title = {Commissioning of a commercial treatment planning system for scanned carbon‐ion radiotherapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  volume = {26},
  number = {3},
  doi = {10.1002/acm2.14580}
}
Start for free →