Long‐term performance monitoring of a‐Si 1200 electronic portal imaging device for dosimetric applications

Ivan Kutuzov iD, Ryan Rivest, Eric VanUytven, Boyd McCurdy

Journal of Applied Clinical Medical Physics · 2025-01 · 26 (1)

DOI: 10.1002/acm2.14551 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 07.10.2024. Verfasst von Ivan Kutuzov und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, Canada. Hauptthemen laut MEDLINE: Particle Accelerators, Radiotherapy Dosage, Radiometry, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Recently, dosimetri applications of the electronic portal imaging device (EPID) in radiotherapy have gained popularity. Confidence in the robust and reliable dosimetric performance of EPID detectors is essential for their clinical use. This study aimed to evaluate the dosimetric performance of the a‐Si 1200 EPID and assess the long‐term stability of its response.

Methods

Weekly measurements were performed on two clinically used TrueBeam linear accelerators (linacs) equipped with a‐Si 1200 EPID detectors over a 2‐year period. They included dark and flood calibration fields, and EPID response to an open field corrected for the long‐term machine output drift measured with the secondary absolute dosimeters: an ion chamber and an ion chamber array. All measurements were performed using five photon beam energies and two imaging modes: continuous and dosimetry. The measurements were analyzed for constancy and the presence of long‐term trends. Comparisons were made between the two linacs for each beam energy. Pixel sensitivity matrices (PSM) were determined semi‐annually and analyzed for long‐term constancy for both treatment machines.

Results

The long‐term variation of the dark and flood field signals, integrated across the EPID plane, over the entire observation period did not exceed 0.17% and 0.79%, respectively. The output‐corrected EPID response showed long‐term variation from 0.28% to 0.36%, depending on beam energy, while the short‐term variation was 0.04%–0.07% for EPID and 0.02%–0.06% for secondary dosimeters. The long‐term variation of secondary dosimeters was 0.2%–0.3%. PSMs were found to be stable to within 1% for 97.8% of pixels and 2% for 100% of pixels.

Conclusion

Techniques to monitor and assess the long‐term performance of the a‐Si 1200 EPID as a dosimeter were developed and implemented using two TrueBeam linacs. The long‐term variation of the EPID response was within clinical tolerance indicated in AAPM TG‐142 report, and the detector was shown to be stable and reproducible for routine clinical dosimetry.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Particle AcceleratorsRadiotherapy DosageRadiometryRadiotherapy Planning, Computer-AssistedHumansCalibrationRadiotherapy, Intensity-ModulatedElectrical Equipment and Supplies

Stichwörter der Autor:innen

EPIDEPID dosimetryEPID response stabilityRadiotherapylinac QAlinac outputmachine QA

Autor:innen und Institutionen

  • Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, Canada.
  • Medical Physics Department, CancerCare Manitoba, Winnipeg, Manitoba, Canada.
  • Department of Radiology, University of Manitoba, Winnipeg, Manitoba, Canada.

ORCID-Profile: Ivan Kutuzov

Zitieren

APA

Kutuzov, I., Rivest, R., VanUytven, E., & McCurdy, B. (2025). Long‐term performance monitoring of a‐Si 1200 electronic portal imaging device for dosimetric applications. Journal of Applied Clinical Medical Physics, 26(1). https://doi.org/10.1002/acm2.14551

Vancouver

Kutuzov I, Rivest R, VanUytven E, McCurdy B. Long‐term performance monitoring of a‐Si 1200 electronic portal imaging device for dosimetric applications. Journal of Applied Clinical Medical Physics. 2025;26(1). doi:10.1002/acm2.14551

BibTeX
@article{Kutuzov2025,
  author = {Kutuzov, Ivan and Rivest, Ryan and VanUytven, Eric and McCurdy, Boyd},
  title = {Long‐term performance monitoring of a‐Si 1200 electronic portal imaging device for dosimetric applications},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  volume = {26},
  number = {1},
  doi = {10.1002/acm2.14551}
}
Kostenlos starten →