Dose rate correction for the novel 2D diode array MapCHECK 3
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14471 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Mengyang Li und 9 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Oncology Center, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences, …. Hauptthemen laut MEDLINE: Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To investigate the dose rate dependence of MapCHECK3 and its influence on measurement accuracy, as well as the effect of dose rate correction. Materials and methods The average and instantaneous dose rate dependence of MapCHECK2 and MapCHECK3 were studied. The accuracy of measurements was investigated where the dose rate differed significantly between dose calibration of the MapCHECK and the measurement. Measurements investigated include: the central axis dose for different fields at different depths, off‐axis doses outside the field, and off‐axis doses along the wedge direction. Measurements using an ion chamber were taken as the reference. Exponential functions were fit to account for average and instantaneous dose rate dependence for MapCHECK3 and used for dose rate correction. The effect of the dose rate correction was studied by comparing the differences between the measurements for MapCHECK (with and without the correction) and the reference.
Results
The maximum dose rate dependence of MapCHECK3 is greater than 2.5%. If the dose calibration factor derived from a 10 × 10 cm 2 open field at 10 cm depth was used for measurements, the average differences in central diode dose were 0.8% ± 1.0% and 1.0% ± 0.8% for the studied field sizes and measurement depths, respectively. The introduction of wedge would not only induce −1.8% ± 1.3% difference in central diode dose, but also overestimate the effective wedge angle. After the instantaneous dose rate correction, above differences can be changed to 1.9% ± 8.1%, 0.2% ± 0.1%, and 0.0% ± 0.9%. The pass rate can be improved from 98.4% to 98.8%, 98.3%–100.0%, and 96.3%–100.0%, respectively.
Conclusion
Compared with MapCHECK2 (SunPoint1 diodes), the more pronounced dose rate dependence of MapCHECK3 (SunPoint2 diodes) should be carefully considered. To ensure highly accurate measurement, it is suggested to perform the dose calibration at the same condition where measurement will be performed. Otherwise, the dose rate correction should be applied.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Radiation Oncology Center, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences, Beijing, China.
- Radiation Oncology Center, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital (Langfang campus), Chinese Academy of Medical Sciences, Langfang, China.
ORCID-Profile: Shouping Xu
Zitieren
APA
Li, M., Tian, Y., Shen, L., Li, G., Zhao, L., Chen, X., Xu, S., Li, M., Huang, P., & Dai, J. (2024). Dose rate correction for the novel 2D diode array MapCHECK 3. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14471
Vancouver
Li M, Tian Y, Shen L, Li G, Zhao L, Chen X, et al. Dose rate correction for the novel 2D diode array MapCHECK 3. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14471
BibTeX
@article{Li2024,
author = {Li, Mengyang and Tian, Yuan and Shen, Linyi and Li, Guiyuan and Zhao, Liang and Chen, Xinyuan and Xu, Shouping and Li, Minghui and Huang, Peng and Dai, Jianrong},
title = {Dose rate correction for the novel 2D diode array MapCHECK 3},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14471}
}
EUVIMED