Dose rate correction for the novel 2D diode array MapCHECK 3

Mengyang Li, Yuan Tian, Linyi Shen, Guiyuan Li, Liang Zhao, Xinyuan Chen, Shouping Xu iD, Minghui Li, Peng Huang, Jianrong Dai

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14471 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Mengyang Li und 9 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Oncology Center, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences, …. Hauptthemen laut MEDLINE: Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To investigate the dose rate dependence of MapCHECK3 and its influence on measurement accuracy, as well as the effect of dose rate correction. Materials and methods The average and instantaneous dose rate dependence of MapCHECK2 and MapCHECK3 were studied. The accuracy of measurements was investigated where the dose rate differed significantly between dose calibration of the MapCHECK and the measurement. Measurements investigated include: the central axis dose for different fields at different depths, off‐axis doses outside the field, and off‐axis doses along the wedge direction. Measurements using an ion chamber were taken as the reference. Exponential functions were fit to account for average and instantaneous dose rate dependence for MapCHECK3 and used for dose rate correction. The effect of the dose rate correction was studied by comparing the differences between the measurements for MapCHECK (with and without the correction) and the reference.

Results

The maximum dose rate dependence of MapCHECK3 is greater than 2.5%. If the dose calibration factor derived from a 10 × 10 cm 2 open field at 10 cm depth was used for measurements, the average differences in central diode dose were 0.8% ± 1.0% and 1.0% ± 0.8% for the studied field sizes and measurement depths, respectively. The introduction of wedge would not only induce −1.8% ± 1.3% difference in central diode dose, but also overestimate the effective wedge angle. After the instantaneous dose rate correction, above differences can be changed to 1.9% ± 8.1%, 0.2% ± 0.1%, and 0.0% ± 0.9%. The pass rate can be improved from 98.4% to 98.8%, 98.3%–100.0%, and 96.3%–100.0%, respectively.

Conclusion

Compared with MapCHECK2 (SunPoint1 diodes), the more pronounced dose rate dependence of MapCHECK3 (SunPoint2 diodes) should be carefully considered. To ensure highly accurate measurement, it is suggested to perform the dose calibration at the same condition where measurement will be performed. Otherwise, the dose rate correction should be applied.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy DosageRadiotherapy Planning, Computer-AssistedHumansCalibrationAlgorithmsRadiotherapy, Intensity-ModulatedRadiometryParticle AcceleratorsPhantoms, Imaging

Stichwörter der Autor:innen

MapCHECK 2MapCHECK 3correctiondose rate dependencedose verification

Autor:innen und Institutionen

  • Radiation Oncology Center, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences, Beijing, China.
  • Radiation Oncology Center, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital (Langfang campus), Chinese Academy of Medical Sciences, Langfang, China.

ORCID-Profile: Shouping Xu

Zitieren

APA

Li, M., Tian, Y., Shen, L., Li, G., Zhao, L., Chen, X., Xu, S., Li, M., Huang, P., & Dai, J. (2024). Dose rate correction for the novel 2D diode array MapCHECK 3. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14471

Vancouver

Li M, Tian Y, Shen L, Li G, Zhao L, Chen X, et al. Dose rate correction for the novel 2D diode array MapCHECK 3. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14471

BibTeX
@article{Li2024,
  author = {Li, Mengyang and Tian, Yuan and Shen, Linyi and Li, Guiyuan and Zhao, Liang and Chen, Xinyuan and Xu, Shouping and Li, Minghui and Huang, Peng and Dai, Jianrong},
  title = {Dose rate correction for the novel 2D diode array MapCHECK 3},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14471}
}
Kostenlos starten →