Beyond conventional bounds: Surpassing system limits for stereotactic ablative (SAbR) lung radiotherapy using CBCT‐based adaptive planning system
Journal of Applied Clinical Medical Physics · 2024-08 · 25 (8)
DOI: 10.1002/acm2.14375 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 07.05.2024. Verfasst von Yesenia Gonzalez und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Lung Neoplasms, Cone-Beam Computed Tomography, Radiotherapy Dosage. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
Purpose
Online adaptive radiotherapy relies on a high degree of automation to enable rapid planning procedures. The Varian Ethos intelligent optimization engine (IOE) was originally designed for conventional treatments so it is crucial to provide clear guidance for lung SAbR plans. This study investigates using the Ethos IOE together with adaptive‐specific optimization tuning structures we designed and templated within Ethos to mitigate inter‐planner variability in meeting RTOG metrics for both online‐adaptive and offline SAbR plans.
Methods
We developed a planning strategy to automate the generation of tuning structures and optimization. This was validated by retrospective analysis of 35 lung SAbR cases (total 105 fractions) treated on Ethos. The effectiveness of our planning strategy was evaluated by comparing plan quality with‐and‐without auto‐generated tuning structures. Internal target volume (ITV) contour was compared between that drawn from CT simulation and from cone‐beam CT (CBCT) at time of treatment to verify CBCT image quality and treatment effectiveness. Planning strategy robustness for lung SAbR was quantified by frequency of plans meeting reference plan RTOG constraints.
Results
Our planning strategy creates a gradient within the ITV with maximum dose in the core and improves intermediate dose conformality on average by 2%. ITV size showed no significant difference between those contoured from CT simulation and first fraction, and also trended towards decreasing over course of treatment. Compared to non‐adaptive plans, adaptive plans better meet reference plan goals (37% vs. 100% PTV coverage compliance, for scheduled and adapted plans) while improving plan quality (improved GI (gradient index) by 3.8%, CI (conformity index) by 1.7%).
Conclusion
We developed a robust and readily shareable planning strategy for the treatment of adaptive lung SAbR on the Ethos system. We validated that automatic online plan re‐optimization along with the formulated adaptive tuning structures can ensure consistent plan quality. With the proposed planning strategy, highly ablative treatments are feasible on Ethos.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
- Department of Radiation Oncology, Memorial Sloan Kettering Cancer Center, New York City, New York, USA.
- Medical Artificial Intelligence and Automation Laboratory, Department of Radiation Oncology, University of Texas Southwestern Medical Center, Dallas, Texas, USA.
ORCID-Profile: David Parsons
Förderung
- NCI NIH HHS
Zitieren
APA
Gonzalez, Y., Visak, J., Overman, L., Liao, C. Y., Yen, A., Zhuang, T., Cai, B., Godley, A., Zhang, Y., Timmerman, R., Iyengar, P., Westover, K., Parsons, D., & Lin, M. H. (2024). Beyond conventional bounds: Surpassing system limits for stereotactic ablative (SAbR) lung radiotherapy using CBCT‐based adaptive planning system. Journal of Applied Clinical Medical Physics, 25(8). https://doi.org/10.1002/acm2.14375
Vancouver
Gonzalez Y, Visak J, Overman L, Liao CY, Yen A, Zhuang T, et al. Beyond conventional bounds: Surpassing system limits for stereotactic ablative (SAbR) lung radiotherapy using CBCT‐based adaptive planning system. Journal of Applied Clinical Medical Physics. 2024;25(8). doi:10.1002/acm2.14375
BibTeX
@article{Gonzalez2024,
author = {Gonzalez, Yesenia and Visak, Justin and Overman, Luke and Liao, Chien-Yi and Yen, Allen and Zhuang, Tingliang and Cai, Bin and Godley, Andrew and Zhang, Yuanyuan and Timmerman, Robert and Iyengar, Puneeth and Westover, Kenneth and Parsons, David and Lin, Mu-Han},
title = {Beyond conventional bounds: Surpassing system limits for stereotactic ablative (SAbR) lung radiotherapy using CBCT‐based adaptive planning system},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
volume = {25},
number = {8},
doi = {10.1002/acm2.14375}
}
EUVIMED