EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Impact of setup errors on the robustness of linac‐based single‐isocenter coplanar and non‐coplanar VMAT plans for multiple brain metastases

Xiaohuan Sun, Fada Guan, Qinghui Yun, Matthew Jennings, Simon Biggs, Zhongfei Wang iD, Wei Wang, Te Zhang, Mei Shi, Lina Zhao

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14317 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Xiaohuan Sun und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Xijing Hospital, Air Force Medical University, Xi'an, China. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Brain Neoplasms, Radiotherapy, Intensity-Modulated, Radiotherapy Dosage. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Patient setup errors have been a primary concern impacting the dose delivery accuracy in radiation therapy. A robust treatment plan might mitigate the effects of patient setup errors. In this reported study, we aimed to evaluate the impact of translational and rotational errors on the robustness of linac‐based, single‐isocenter, coplanar, and non‐coplanar volumetric modulated arc therapy treatment plans for multiple brain metastases.

Methods

Fifteen patients were retrospectively selected for this study with a combined total of 49 gross tumor volumes (GTVs). Single‐isocenter coplanar and non‐coplanar plans were generated first with a prescribed dose of 40 Gy in 5 fractions or 42 Gy in 7 fractions to cover 95% of planning target volume (PTV). Next, four setup errors (+1 and +2 mm translation, and +1° and +2° rotation) were applied individually to generate modified plans. Different plan quality evaluation metrics were compared between coplanar and non‐coplanar plans. 3D gamma analysis (3%/2 mm) was performed to compare the modified plans (+2 mm and +2° only) and the original plans. Paired t ‐test was conducted for statistical analysis.

Results

After applying setup errors, variations of all plan evaluation metrics were similar ( p > 0.05). The worst case for V100% to GTV was 92.07% ± 6.13% in the case of +2 mm translational error. 3D gamma pass rates were > 90% for both coplanar (+2 mm and +2°) and the +2 mm non‐coplanar groups but was 87.40% ± 6.89% for the +2° non‐coplanar group.

Conclusion

Translational errors have a greater impact on PTV and GTV dose coverage for both planning methods. Rotational errors have a greater negative impact on gamma pass rates of non‐coplanar plans. Plan evaluation metrics after applying setup errors showed that both coplanar and non‐coplanar plans were robust and clinically acceptable.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedBrain NeoplasmsRadiotherapy, Intensity-ModulatedRadiotherapy DosageRadiotherapy Setup ErrorsParticle AcceleratorsOrgans at RiskHumansRetrospective StudiesPrognosisPatient Positioning

Stichwörter der Autor:innen

coplanar and noncoplanar VMAT plansmultiple brain metastasespatient setup errorssingle isocenter

Autor:innen und Institutionen

  • Department of Radiation Oncology, Xijing Hospital, Air Force Medical University, Xi'an, China.
  • Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut, USA.
  • Department of Equipment, Xijing Hospital, Air Force Medical University, Xi'an, China.
  • Department of Medical Physics, Townsville University Hospital, Douglas, Queensland, Australia.
  • Radiotherapy AI Pty Ltd, Wagga Wagga, Australia.

ORCID-Profile: Zhongfei Wang

Förderung

  • National Natural Science Foundation of China

Zitieren

APA

Sun, X., Guan, F., Yun, Q., Jennings, M., Biggs, S., Wang, Z., Wang, W., Zhang, T., Shi, M., & Zhao, L. (2024). Impact of setup errors on the robustness of linac‐based single‐isocenter coplanar and non‐coplanar VMAT plans for multiple brain metastases. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14317

Vancouver

Sun X, Guan F, Yun Q, Jennings M, Biggs S, Wang Z, et al. Impact of setup errors on the robustness of linac‐based single‐isocenter coplanar and non‐coplanar VMAT plans for multiple brain metastases. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14317

BibTeX
@article{Sun2024,
  author = {Sun, Xiaohuan and Guan, Fada and Yun, Qinghui and Jennings, Matthew and Biggs, Simon and Wang, Zhongfei and Wang, Wei and Zhang, Te and Shi, Mei and Zhao, Lina},
  title = {Impact of setup errors on the robustness of linac‐based single‐isocenter coplanar and non‐coplanar VMAT plans for multiple brain metastases},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14317}
}