Independent Monte Carlo dose calculation identifies single isocenter multi‐target radiosurgery targets most likely to fail pre‐treatment measurement

Brett Erickson, Yunfeng Cui, Markus Alber, Chunhao Wang, Fang Fang Yin, John Kirkpatrick, Justus Adamson

Journal of Applied Clinical Medical Physics · 2024-06 · 25 (6)

DOI: 10.1002/acm2.14290 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 30.01.2024. Verfasst von Brett Erickson und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina, USA. Hauptthemen laut MEDLINE: Monte Carlo Method, Radiosurgery, Radiotherapy Planning, Computer-Assisted, Radiotherapy Dosage. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

For individual targets of single isocenter multi‐target (SIMT) Stereotactic radiosurgery (SRS), we assess dose difference between the treatment planning system (TPS) and independent Monte Carlo (MC), and demonstrate persistence into the pre‐treatment Quality Assurance (QA) measurement.

Methods

Treatment plans from 31 SIMT SRS patients were recalculated in a series of scenarios designed to investigate sources of discrepancy between TPS and independent MC. Targets with > 5% discrepancy in DMean[Gy] after progressing through all scenarios were measured with SRS MapCHECK. A matched pair analysis was performed comparing SRS MapCHECK results for these targets with matched targets having similar characteristics (volume & distance from isocenter) but no such MC dose discrepancy.

Results

Of 217 targets analyzed, individual target mean dose (DMean[Gy]) fell outside a 5% threshold for 28 and 24 targets before and after removing tissue heterogeneity effects, respectively, while only 5 exceeded the threshold after removing effect of patient geometry (via calculation on StereoPHAN geometry). Significant factors affecting agreement between the TPS and MC included target distance from isocenter (0.83% decrease in DMean[Gy] per 2 cm), volume (0.15% increase per cc), and degree of plan modulation (0.37% increase per 0.01 increase in modulation complexity score). SRS MapCHECK measurement had better agreement with MC than with TPS (2%/1 mm / 10% threshold gamma pass rate (GPR) = 99.4 ± 1.9% vs. 93.1 ± 13.9%, respectively). In the matched pair analysis, targets exceeding 5% for MC versus TPS also had larger discrepancies between TPS and measurement with no GPR (2%/1 mm / 10% threshold) exceeding 90% (71.5% ± 16.1%); whereas GPR was high for matched targets with no such MC versus TPS difference (96.5% ± 3.3%, p = 0.01).

Conclusions

Independent MC complements pre‐treatment QA measurement for SIMT SRS by identifying problematic individual targets prior to pre‐treatment measurement, thus enabling plan modifications earlier in the planning process and guiding selection of targets for pre‐treatment QA measurement.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodRadiosurgeryRadiotherapy Planning, Computer-AssistedRadiotherapy DosageRadiotherapy, Intensity-ModulatedQuality Assurance, Health CareHumansOrgans at RiskAlgorithmsNeoplasms

Stichwörter der Autor:innen

dosimetric robustnessquality assurancesingle isocenter multi‐target radiosurgery

Autor:innen und Institutionen

  • Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina, USA.
  • Scientific RT, Munich, Germany.

Zitieren

APA

Erickson, B., Cui, Y., Alber, M., Wang, C., Fang Yin, F., Kirkpatrick, J., & Adamson, J. (2024). Independent Monte Carlo dose calculation identifies single isocenter multi‐target radiosurgery targets most likely to fail pre‐treatment measurement. Journal of Applied Clinical Medical Physics, 25(6). https://doi.org/10.1002/acm2.14290

Vancouver

Erickson B, Cui Y, Alber M, Wang C, Fang Yin F, Kirkpatrick J, et al. Independent Monte Carlo dose calculation identifies single isocenter multi‐target radiosurgery targets most likely to fail pre‐treatment measurement. Journal of Applied Clinical Medical Physics. 2024;25(6). doi:10.1002/acm2.14290

BibTeX
@article{Erickson2024,
  author = {Erickson, Brett and Cui, Yunfeng and Alber, Markus and Wang, Chunhao and Fang Yin, Fang and Kirkpatrick, John and Adamson, Justus},
  title = {Independent Monte Carlo dose calculation identifies single isocenter multi‐target radiosurgery targets most likely to fail pre‐treatment measurement},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  volume = {25},
  number = {6},
  doi = {10.1002/acm2.14290}
}
Kostenlos starten →