Assessment of dose distribution in pancreatic SABR in the presence of a metallic biliary stent using Monte Carlo modeling

Ahmad Abbas iD, Owen Nicholas, Richard P Hugtenburg

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70712 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Ahmad Abbas und 2 weiteren Autor:innen. Beteiligte Einrichtung: Swansea University Medical School, Swansea University, Swansea, Wales, UK. Hauptthemen laut MEDLINE: Monte Carlo Method, Pancreatic Neoplasms, Radiosurgery, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Stereotactic ablative radiotherapy (SABR) with dose escalation is increasingly used for pancreatic cancer; however, the dosimetric impact of metallic biliary stents remains poorly characterized. Self‐expanding metal stents (SEMS), commonly placed for obstructive jaundice, contain high‐density nitinol that produces computed tomography (CT) artifacts, including beam hardening, potentially compromising treatment planning system (TPS) dose accuracy. Precise dose estimation is critical given the proximity of the duodenum and other organs at risk (OARs).

Purpose

This study evaluated dose distribution in pancreatic SABR in the presence of a metallic biliary stent using Monte Carlo (MC) simulation with refined stent contouring and material correction.

Methods

Two patients with SEMS (Cook Evolution, nitinol alloy) treated with SABR (50 Gy in 5 fractions) using volumetric modulated arc therapy (VMAT) were retrospectively analyzed. Plans were generated in Eclipse TPS (v.13.7) using the analytical anisotropic algorithm. MC simulations were performed in PRIMO (PENELOPE‐based) with ≤2% uncertainty. A novel contouring technique separated the hollow lumen from the metallic stent wall to correct CT beam hardening artifacts. Nitinol (density 6.45 g/cm 3 ) was incorporated into MC calibration. Comparisons between standard MC, modified MC, and TPS were assessed using percentage agreement (PA), gamma passing rate (GPR; 3%/3 mm), and dose‐volume histogram metrics.

Results

Refined contouring reduced modeled stent volumes from 26.1 to 22.8 cm 3 and from 19.87 to 18.23 cm 3 . The standard MC model overestimated planning target volume (PTV) dose by 14.9% and 15.7%, with PTV GPRs of 55.29% and 79.78%, respectively. Approximately 2% dose enhancement was observed at the stent‐tissue interface. Compared with the modified MC model, TPS underestimated dose lateral to the stent by up to 12.3% (worst case, Patient B, Plan 2) and overestimated it elsewhere by up to 12.2%. The duodenum showed the largest OAR discrepancies, with TPS up to 9.7% underestimation adjacent to the stent, whereas distant structures demonstrated high agreement.

Conclusion

Conventional TPS algorithms exhibit clinically relevant inaccuracies in the presence of metallic biliary stents. Incorporating refined stent contouring and material correction within an MC framework improves dosimetric accuracy and may support safer dose escalation in pancreatic SABR.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monte Carlo MethodPancreatic NeoplasmsRadiosurgeryRadiotherapy Planning, Computer-AssistedStentsHumansRadiotherapy DosageOrgans at RiskRadiotherapy, Intensity-ModulatedRetrospective StudiesAlloysAlgorithmsPrognosis

Stichwörter der Autor:innen

Monte Carlo simulationPRIMOVMATbiliary stentdose distributionmetal artifactnitinolpancreatic cancerstereotactic ablative radiotherapy (SABR)

Substanzen

Alloysnitinol

Autor:innen und Institutionen

  • Swansea University Medical School, Swansea University, Swansea, Wales, UK.
  • Adan Hospital, Ministry of Health, Hadiya, Kuwait.
  • South West Wales Cancer Centre, Swansea Bay University Health Board, Swansea, Wales, UK.

ORCID-Profile: Ahmad Abbas

Zitieren

APA

Abbas, A., Nicholas, O., & Hugtenburg, R. P. (2026). Assessment of dose distribution in pancreatic SABR in the presence of a metallic biliary stent using Monte Carlo modeling. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70712

Vancouver

Abbas A, Nicholas O, Hugtenburg RP. Assessment of dose distribution in pancreatic SABR in the presence of a metallic biliary stent using Monte Carlo modeling. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70712

BibTeX
@article{Abbas2026,
  author = {Abbas, Ahmad and Nicholas, Owen and Hugtenburg, Richard P},
  title = {Assessment of dose distribution in pancreatic SABR in the presence of a metallic biliary stent using Monte Carlo modeling},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70712}
}
Kostenlos starten →