Influence of magnetic field on a novel scintillation dosimeter in a 1.5 T MR‐linac

Stijn Oolbekkink iD, Bram van Asselen, Simon J Woodings, Jochem W H Wolthaus, J H Wilfred de Vries, Adriaan A van Appeldoorn, Marcos Feijoo, Madelon van den Dobbelsteen iD, Bas W Raaymakers

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14180 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Stijn Oolbekkink und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiotherapy, University Medical Center Utrecht, Utrecht, The Netherlands. Hauptthemen laut MEDLINE: Radiation Dosimeters, Particle Accelerators. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Nederlandse Organisatie voor Wetenschappelijk Onderzoek, European Metrology Programme for Innovation and Research.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

For commissioning and quality assurance for adaptive workflows on the MR‐linac, a dosimeter which can measure time‐resolved dose during MR image acquisition is desired. The Blue Physics model 10 scintillation dosimeter is potentially an ideal detector for such measurements. However, some detectors can be influenced by the magnetic field of the MR‐linac. To assess the calibration methods and magnetic field dependency of the Blue Physics scintillator in the 1.5 T MR‐linac. Several calibration methods were assessed for robustness. Detector characteristics and the influence of the calibration methods were assessed based on dose reproducibility, dose linearity, dose rate dependency, relative output factor (ROF), percentage depth dose profile, axial rotation and the radial detector orientation with respect to the magnetic field. The potential application of time‐resolved dynamic dose measurements during MRI acquisition was assessed. A variation of calibration factors was observed for different calibration methods. Dose reproducibility, dose linearity and dose rate stability were all found to be within tolerance and were not significantly affected by different calibration methods. Measurements with the detector showed good correspondence with reference chambers. The ROF and radial orientation dependence measurements were influenced by the calibration method used. Axial detector dependence was assessed and relative readout differences of up to 2.5% were observed. A maximum readout difference of 10.8% was obtained when rotating the detector with respect to the magnetic field. Importantly, measurements with and without MR image acquisition were consistent for both static and dynamic situations. The Blue Physics scintillation detector is suitable for relative dosimetry in the 1.5 T MR‐linac when measurements are within or close to calibration conditions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosimetersParticle AcceleratorsHumansReproducibility of ResultsPhantoms, ImagingRadiometryMagnetic Resonance ImagingMagnetic Fields

Stichwörter der Autor:innen

MRI-linacradiotherapyscintillation dosimetrytime-resolved dosimetry

Autor:innen und Institutionen

  • Department of Radiotherapy, University Medical Center Utrecht, Utrecht, The Netherlands.
  • Blue Physics, Lutz, Florida, USA.

ORCID-Profile: Stijn Oolbekkink, Madelon van den Dobbelsteen

Förderung

  • Nederlandse Organisatie voor Wetenschappelijk Onderzoek
  • European Metrology Programme for Innovation and Research

Zitieren

APA

Oolbekkink, S., van Asselen, B., Woodings, S. J., Wolthaus, J. W. H., de Vries, J. H. W., van Appeldoorn, A. A., Feijoo, M., van den Dobbelsteen, M., & Raaymakers, B. W. (2024). Influence of magnetic field on a novel scintillation dosimeter in a 1.5 T MR‐linac. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14180

Vancouver

Oolbekkink S, van Asselen B, Woodings SJ, Wolthaus JWH, de Vries JHW, van Appeldoorn AA, et al. Influence of magnetic field on a novel scintillation dosimeter in a 1.5 T MR‐linac. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14180

BibTeX
@article{Oolbekkink2024,
  author = {Oolbekkink, Stijn and van Asselen, Bram and Woodings, Simon J and Wolthaus, Jochem W H and de Vries, J H Wilfred and van Appeldoorn, Adriaan A and Feijoo, Marcos and van den Dobbelsteen, Madelon and Raaymakers, Bas W},
  title = {Influence of magnetic field on a novel scintillation dosimeter in a 1.5 T MR‐linac},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14180}
}
Kostenlos starten →