Magnetic field quality conversion factors experimentally measured in clinical MR‐linac beams for seven MR‐compatible ionization chamber models
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.14613 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Nathan Orlando und 4 weiteren Autor:innen. Beteiligte Einrichtung: Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario, Canada. Hauptthemen laut MEDLINE: Particle Accelerators, Magnetic Fields, Magnetic Resonance Imaging, Radiometry. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NHLBI NIH HHS, National Cancer Institute of the National Institutes of Health, Natural Sciences and Engineering Research Council.
Abstract (Originalsprache)
Purpose
The purpose of this work was to experimentally quantify MR‐compatible ionization chamber response for 1.5T Elekta Unity and 0.35T ViewRay MRIdian MR‐linac systems through the determination of the magnetic field quality conversion factor, k B,Q .
Methods
Seven MR‐compatible ionization chamber models from Standard Imaging and PTW were evaluated. Both the quality conversion factor k Q and the magnetic field quality conversion factor k B,Q were experimentally determined through a cross‐calibration method. Specifically, the ratio of absorbed dose measured with a reference A1SL chamber under reference conditions to corrected output measured with each test chamber at the same point of measurement allowed for the determination of k B,Q . The angular dependence of the magnetic field quality conversion factor for MR‐compatible chamber models was assessed for the 1.5T Elekta Unity system by measuring k B,Q with the chamber axis and magnetic field direction aligned at cardinal angles (0°, 90°, 180°, 270°).
Results
Beam quality conversion ( k Q ) factors for MR‐compatible ionization chambers measured in a standard linac beam showed an average percent difference of −0.09 ± 0.18% compared to computed k Q values for their conventional chamber versions. Similarly, magnetic field quality conversion ( k B,Q ) factors for corresponding MR and non‐MR ionization chamber models measured using the same cross‐calibration technique demonstrated average percent differences of −0.1 ± 0.3% and 0.0 ± 0.2% for the Elekta Unity and ViewRay MRIdian, respectively. Investigation of the angular dependence of this correction factor demonstrated identical chamber response for equivalent MR‐compatible and conventional chamber models.
Conclusions
This work provides critical experimental validation of MR‐compatible ionization chamber performance, with a direct comparison of measured k B,Q values to corresponding conventional chamber models demonstrating nearly equivalent chamber response. k B,Q values determined using our experimental method will serve as an important reference for upcoming MR‐linac reference dosimetry protocols and ultimately represent an important step towards accurate output calibration of MR‐linac systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario, Canada.
- Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada.
- Department of Human Oncology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Förderung
- NHLBI NIH HHS
- National Cancer Institute of the National Institutes of Health
- Natural Sciences and Engineering Research Council
Zitieren
APA
Orlando, N., Crosby, J., Glide-Hurst, C., Culberson, W., & Sarfehnia, A. (2025). Magnetic field quality conversion factors experimentally measured in clinical MR‐linac beams for seven MR‐compatible ionization chamber models. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14613
Vancouver
Orlando N, Crosby J, Glide-Hurst C, Culberson W, Sarfehnia A. Magnetic field quality conversion factors experimentally measured in clinical MR‐linac beams for seven MR‐compatible ionization chamber models. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.14613
BibTeX
@article{Orlando2025,
author = {Orlando, Nathan and Crosby, Jennie and Glide-Hurst, Carri and Culberson, Wesley and Sarfehnia, Arman},
title = {Magnetic field quality conversion factors experimentally measured in clinical MR‐linac beams for seven MR‐compatible ionization chamber models},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.14613}
}
EUVIMED