Impact of magnetic fields on small field dosimetry for parallel MR‐linacs
Medical Physics · 2025-08 · 52 (8)
DOI: 10.1002/mp.18052 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, online erschienen am 13.08.2025. Verfasst von Ahtesham Ullah Khan. Beteiligte Einrichtung: Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Magnetic Fields, Radiometry, Magnetic Resonance Imaging, Particle Accelerators. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Science Foundation.
Abstract (Originalsprache)
Background
The use of MR‐guided radiation therapy (MRgRT) is increasing with the introduction of commercially‐available parallel MR‐linac systems. Treatment fields used by these machines can be very small and highly modulated. Literature on the small field dosimetry of parallel MR‐linacs and the impact of the magnetic field on dose perturbations is scarce.
Purpose
To investigate the impact of magnetic fields on small field dosimetry of parallel MR‐linacs using Monte Carlo (MC) methods.
Methods
A TOPAS MC model of a 6 MV FFF beam was developed and validated against a commercial 0.5 T parallel MR‐linac. The impact of varying magnetic field of up to 3.0 T on water phantom dosimetry such as percent depth dose (PDD), beam profiles, and output factors (OFs) was studied. A lung phantom with embedded spherical tumors of diameters 1–3 cm was employed to investigate the impact of parallel magnetic fields on lung lesions using an arc delivery with a 4 ο gantry spacing providing a surrogate for rotational delivery techniques. Dose distributions were compared with the 0.0 T scenario.
Results
Negligible differences were noted in the PDD, except for the build‐up region, between the investigated parallel magnetic fields. Besides the tail region of the 1.5 and 3.0 T profiles, minimal differences were observed in the lateral beam profiles. Similarly, parallel magnetic fields of up to 1 T had a negligible impact on small field OFs. Compared to the 0 T magnetic field, the OF for field sizes of
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Förderung
- National Science Foundation
Zitieren
APA
Khan, A. U. (2025). Impact of magnetic fields on small field dosimetry for parallel MR‐linacs. Medical Physics, 52(8). https://doi.org/10.1002/mp.18052
Vancouver
Khan AU. Impact of magnetic fields on small field dosimetry for parallel MR‐linacs. Medical Physics. 2025;52(8). doi:10.1002/mp.18052
BibTeX
@article{Khan2025,
author = {Khan, Ahtesham Ullah},
title = {Impact of magnetic fields on small field dosimetry for parallel MR‐linacs},
journal = {Medical Physics},
year = {2025},
volume = {52},
number = {8},
doi = {10.1002/mp.18052}
}
EUVIMED