Impact of magnetic fields on small field dosimetry for parallel MR‐linacs

Ahtesham Ullah Khan

Medical Physics · 2025-08 · 52 (8)

DOI: 10.1002/mp.18052 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Medical Physics, online erschienen am 13.08.2025. Verfasst von Ahtesham Ullah Khan. Beteiligte Einrichtung: Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA. Hauptthemen laut MEDLINE: Magnetic Fields, Radiometry, Magnetic Resonance Imaging, Particle Accelerators. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Science Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The use of MR‐guided radiation therapy (MRgRT) is increasing with the introduction of commercially‐available parallel MR‐linac systems. Treatment fields used by these machines can be very small and highly modulated. Literature on the small field dosimetry of parallel MR‐linacs and the impact of the magnetic field on dose perturbations is scarce.

Purpose

To investigate the impact of magnetic fields on small field dosimetry of parallel MR‐linacs using Monte Carlo (MC) methods.

Methods

A TOPAS MC model of a 6 MV FFF beam was developed and validated against a commercial 0.5 T parallel MR‐linac. The impact of varying magnetic field of up to 3.0 T on water phantom dosimetry such as percent depth dose (PDD), beam profiles, and output factors (OFs) was studied. A lung phantom with embedded spherical tumors of diameters 1–3 cm was employed to investigate the impact of parallel magnetic fields on lung lesions using an arc delivery with a 4 ο gantry spacing providing a surrogate for rotational delivery techniques. Dose distributions were compared with the 0.0 T scenario.

Results

Negligible differences were noted in the PDD, except for the build‐up region, between the investigated parallel magnetic fields. Besides the tail region of the 1.5 and 3.0 T profiles, minimal differences were observed in the lateral beam profiles. Similarly, parallel magnetic fields of up to 1 T had a negligible impact on small field OFs. Compared to the 0 T magnetic field, the OF for field sizes of

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic FieldsRadiometryMagnetic Resonance ImagingParticle AcceleratorsRadiotherapy, Image-GuidedMonte Carlo MethodPhantoms, ImagingRadiotherapy DosageHumans

Stichwörter der Autor:innen

MR‐guided RTMR‐linacsMonte Carlomagnetic fieldssmall field dosimetry

Autor:innen und Institutionen

  • Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, Wisconsin, USA.

Förderung

  • National Science Foundation

Zitieren

APA

Khan, A. U. (2025). Impact of magnetic fields on small field dosimetry for parallel MR‐linacs. Medical Physics, 52(8). https://doi.org/10.1002/mp.18052

Vancouver

Khan AU. Impact of magnetic fields on small field dosimetry for parallel MR‐linacs. Medical Physics. 2025;52(8). doi:10.1002/mp.18052

BibTeX
@article{Khan2025,
  author = {Khan, Ahtesham Ullah},
  title = {Impact of magnetic fields on small field dosimetry for parallel MR‐linacs},
  journal = {Medical Physics},
  year = {2025},
  volume = {52},
  number = {8},
  doi = {10.1002/mp.18052}
}
Kostenlos starten →