Recombination and polarity effects of Farmer chambers in a strong magnetic field

Hans Lynggaard Riis, Kenni Højsgaard Engstrøm, Claus E Andersen

Physica Medica · 2024

DOI: 10.1016/j.ejmp.2024.103406 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physica Medica, 2024. Verfasst von Hans Lynggaard Riis und 2 weiteren Autor:innen. Beteiligte Einrichtung: Odense University Hospital, Department of Oncology, Odense, Denmark; University of Southern Denmark, Department of Clinical Research, Odense, Denmark…. Hauptthemen laut MEDLINE: Magnetic Fields, Radiometry. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: Ionisation chamber based reference dosimetry in magnetic resonance linear accelerators (MRL) aimed for radiotherapy requires correction for recombination losses. Published studies have found that such corrections can be carried out using the two-voltage method. These studies have, however, not included comparison with recombination corrections based on the Niatel method, which can be seen as a robust reference method due to its clear separation of initial and volume recombination and its explicit account of the pulsed nature of the dose delivery. The primary objective of this work therefore was to carry out such a comparison. MATERIALS AND METHODS: Four Farmer-type chambers (PTW-30006 and PTW-30013) were placed in a water phantom in 1.5 T Elekta Unity MRL. The chambers were oriented antiparallel or perpendicular to the static magnetic field B0 and irradiated at a source-to-surface distance of 133.5 cm with a 10 × 10 cm2 field size. RESULTS: The two-voltage method gave results in agreement (within 0.1%) with the recombination corrections derived from the Niatel method. The recombination corrections from three Niatel parameter sets (one based on a Varian Truebeam and two obtained directly in the MRL) deviated less than 0.1% from each other. A systematic shift in the recombination correction of less than 0.05% was observed if polarity corrections were not applied. CONCLUSIONS: The study supports the use of the two-voltage method in MRLs based on its excellent agreement with the Niatel method. This work, therefore, complements existing knowledge as previous studies have not included a comparison with the Niatel method.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Magnetic FieldsRadiometryParticle AcceleratorsPhantoms, Imaging

Stichwörter der Autor:innen

DosimetryIon chambersJaffé plotsMagnetic fieldNiatelPolarity effectsRecombination

Autor:innen und Institutionen

  • Odense University Hospital, Department of Oncology, Odense, Denmark; University of Southern Denmark, Department of Clinical Research, Odense, Denmark. Electronic address: hans.lynggaard.riis@rsyd.dk.
  • Odense University Hospital, Department of Oncology, Odense, Denmark.
  • Technical University of Denmark, Department of Health Technology, Roskilde, Denmark.

Zitieren

APA

Riis, H. L., Engstrøm, K. H., & Andersen, C. E. (2024). Recombination and polarity effects of Farmer chambers in a strong magnetic field. Physica Medica. https://doi.org/10.1016/j.ejmp.2024.103406

Vancouver

Riis HL, Engstrøm KH, Andersen CE. Recombination and polarity effects of Farmer chambers in a strong magnetic field. Physica Medica. 2024. doi:10.1016/j.ejmp.2024.103406

BibTeX
@article{Riis2024,
  author = {Riis, Hans Lynggaard and Engstrøm, Kenni Højsgaard and Andersen, Claus E},
  title = {Recombination and polarity effects of Farmer chambers in a strong magnetic field},
  journal = {Physica Medica},
  year = {2024},
  doi = {10.1016/j.ejmp.2024.103406}
}
Kostenlos starten →