Verification of a Benchtop Hematology Analyzer with a 5-Part Differential Count: There is Nothing Wrong with Being Small

Anima Baiden, Elise Schapkaitz, Maynolia Naidoo, Susan Louw

Clinical Laboratory · 2024

DOI: 10.7754/clin.lab.2023.230630 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinical Laboratory, 2024. Verfasst von Anima Baiden und 3 weiteren Autor:innen. Hauptthemen laut MEDLINE: Hematology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The benchtop ADVIA 560 AL hematology analyzer (Siemens Healthineers Tarrytown, NY, USA) offers a small footprint and ease of operation making it suitable for satellite laboratories and intensive care units. A verification study of this analyzer was performed. METHODS: Between- and intra-run precision, carry-over, linearity, and throughput were evaluated on the ADVIA 560 AL. Accuracy was assessed on 94 patient samples by comparing the results obtained on the ADVIA 560 AL to the results on the reference Sysmex XN1000 analyzer (Sysmex Corporation, Kobe, Japan). RESULTS: The ADVIA 560 AL showed acceptable imprecision on control material and minimal bias in comparison to the XN 1000 on patient samples with a throughput of 60 samples per hour. The percentage carryover was not significant and the linearity was within acceptable limits. CONCLUSIONS: The ADVIA 560 AL bench-top analyzer is suitable for acute care centers and satellite laboratories owing to its small footprint, ease of use, and reproducible and accurate results.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HematologyHumansBlood Cell CountReproducibility of ResultsLaboratoriesJapanLeukocyte Count

Zitieren

APA

Baiden, A., Schapkaitz, E., Naidoo, M., & Louw, S. (2024). Verification of a Benchtop Hematology Analyzer with a 5-Part Differential Count: There is Nothing Wrong with Being Small. Clinical Laboratory. https://doi.org/10.7754/clin.lab.2023.230630

Vancouver

Baiden A, Schapkaitz E, Naidoo M, Louw S. Verification of a Benchtop Hematology Analyzer with a 5-Part Differential Count: There is Nothing Wrong with Being Small. Clinical Laboratory. 2024. doi:10.7754/clin.lab.2023.230630

BibTeX
@article{Baiden2024,
  author = {Baiden, Anima and Schapkaitz, Elise and Naidoo, Maynolia and Louw, Susan},
  title = {Verification of a Benchtop Hematology Analyzer with a 5-Part Differential Count: There is Nothing Wrong with Being Small},
  journal = {Clinical Laboratory},
  year = {2024},
  doi = {10.7754/clin.lab.2023.230630}
}
Kostenlos starten →