A comparative study of blood cell count in four automated hematology analyzers: An evaluation of the impact of preanalytical factors
PLOS ONE · 24.05.2024 · 19 (5) · e0301845
DOI: 10.1371/journal.pone.0301845 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift PLOS ONE, online erschienen am 24.05.2024. Verfasst von Annika Åstrand und 13 weiteren Autor:innen. Beteiligte Einrichtung: Clinical Development, Late-stage Development, Respiratory and Immunology, BioPharmaceuticals R&D, AstraZeneca, Gothenburg, Sweden. Hauptthemen laut MEDLINE: Asthma, Eosinophils. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Differential white blood cell counts are frequently used in diagnosis, patient stratification, and treatment selection to optimize therapy responses. Referral laboratories are often used but challenged with use of different hematology platforms, variable blood shipping times and storage conditions, and the different sensitivities of specific cell types. To extend the scientific literature and knowledge on the temporal commutability of blood samples between hematology analyzers, we performed a comparative ex-vivo study using four of the most utilized commercial platforms, focusing on the assessment of eosinophils given its importance in asthma management. Whole blood from healthy volunteers with and without atopy (n = 6+6) and participants with eosinophilic asthma (n = 6) were stored under different conditions (at 4, 20, 30, and 37°C, with or without agitation) and analyzed at different time points (3, 6, 24, 48 and 72h post-sampling) in parallel on the Abbott CELL-DYN Sapphire, Beckman Coulter DxH900, Siemens ADVIA 2120i and Sysmex XN-1000V. In the same blood samples, eosinophil-derived neurotoxin (EDN), eosinophil activation and death markers were analyzed. All platforms gave comparable measurements of cell differentials on fresh blood within the same day of sampling. However, by 24 hours, significant temporal and temperature-dependent differences were observed, most markedly for eosinophils. None of the platforms performed perfectly across all temperatures tested during the 72 hours, showing that handling conditions should be optimized depending on the cell type of interest and the hematology analyzer. Neither disease status (healthy vs. asthma) nor agitation of the sample affected the cell quantification result or EDN release. The eosinophil activation markers measured by flow cytometry increased with time, were influenced by temperature, and were higher in those with asthma versus healthy participants. In conclusion, hematology analyzer, time window from sampling until analysis, and temperature conditions must be considered when analyzing blood cell differentials, particularly for eosinophils, via central labs to obtain counts comparable to the values obtained in freshly sampled blood.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Clinical Development, Late-stage Development, Respiratory and Immunology, BioPharmaceuticals R&D, AstraZeneca, Gothenburg, Sweden.
- Bioscience, Research and Early Development, Respiratory & Immunology, BioPharmaceuticals R&D, AstraZeneca, Gothenburg, Sweden.
- Biometrics, Late-stage Development, Respiratory and Immunology, BioPharmaceuticals R&D, AstraZeneca, Cambridge, United Kingdom.
- Translational Science & Experimental Medicine, Research and Early Development, Respiratory & Immunology, BioPharmaceuticals R&D, AstraZeneca, Gothenburg, Sweden.
- Integrated Bioanalysis, Clinical Pharmacology and Safety Sciences, R&D, AstraZeneca, Gaithersburg, MD, United States of America.
- Precision Medicine Unit, Biopharmaceuticals R&D, AstraZeneca, Cambridge, United Kingdom.
ORCID-Profile: Annika Åstrand, Björn Burmeister, Gabriela Luporini Saraiva
Zitieren
APA
Åstrand, A., Wingren, C., Walton, C., Mattsson, J., Agrawal, K., Lindqvist, M., Odqvist, L., Burmeister, B., Eck, S., Hughes, G., Luporini Saraiva, G., Schantz, A., Psallidas, I., & McCrae, C. (2024). A comparative study of blood cell count in four automated hematology analyzers: An evaluation of the impact of preanalytical factors. PLOS ONE, 19(5), e0301845. https://doi.org/10.1371/journal.pone.0301845
Vancouver
Åstrand A, Wingren C, Walton C, Mattsson J, Agrawal K, Lindqvist M, et al. A comparative study of blood cell count in four automated hematology analyzers: An evaluation of the impact of preanalytical factors. PLOS ONE. 2024;19(5):e0301845. doi:10.1371/journal.pone.0301845
BibTeX
@article{strand2024,
author = {Åstrand, Annika and Wingren, Cecilia and Walton, Claire and Mattsson, Johan and Agrawal, Komal and Lindqvist, Madelene and Odqvist, Lina and Burmeister, Björn and Eck, Steven and Hughes, Glen and Luporini Saraiva, Gabriela and Schantz, Anna and Psallidas, Ioannis and McCrae, Christopher},
title = {A comparative study of blood cell count in four automated hematology analyzers: An evaluation of the impact of preanalytical factors},
journal = {PLOS ONE},
year = {2024},
volume = {19},
number = {5},
pages = {e0301845},
doi = {10.1371/journal.pone.0301845}
}
EUVIMED