Comparative Evaluation of Four Hematology Analyzers: Siemens Advia 2120i, Sysmex XN ‐1000, Mindray BC ‐5310 CRP , and Beckman Coulter DxH 900. Do Similar Performances Equal Same Quality?

Ana Nikler iD, Marija Grdić Rajković, Vanja Radišić Biljak

International Journal of Laboratory Hematology · 2025

DOI: 10.1111/ijlh.70025 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift International Journal of Laboratory Hematology, 2025. Verfasst von Ana Nikler und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Laboratory Diagnostics, University Hospital "Sveti Duh", Zagreb, Croatia. Hauptthemen laut MEDLINE: Hematology. Die Arbeit nennt 25 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

This study provides a comprehensive verification and comparative evaluation of four hematology analyzers (HA) with a five‐part differential blood count (5‐DIFF) (Siemens Advia 2120i, Sysmex XN‐1000, Mindray BC‐5310CRP, Beckman Coulter DxH 900).

Methods

The parameters of complete blood count (CBC) with 5‐DIFF were examined. The verification protocol assessed precision, accuracy, linearity, limits of blank (LoB), detection (LoD) and quantification (LoQ), and carryover. Advia 2120i served as the reference HA for accuracy assessment. Statistical analysis was performed using MedCalc and MS Excel. Acceptance criteria followed manufacturer specifications and biological variability analytical performance specifications (APS).

Results

All analyzers showed acceptable precision for most CBC parameters, while XN‐1000 and DxH 900 showed the best repeatability. BC‐5310 CRP showed the highest proportion of parameters exceeding acceptance criteria: low leukocytes in control (2.9%), and low platelets (6.1%) as well as low (16.4%) and high (4.1%) leukocytes in patient samples. Advia 2120i showed suboptimal repeatability at low hemoglobin in control (53 g/L, CV 1.1%) and low leukocytes (4.7%) for patient samples. Platelet repeatability varied most (CV 5.4%–9.7%). Platelet accuracy was significantly different on XN‐1000 ( p = 0.04) and BC‐5310CRP ( p = 0.05). Determined LoQ for leukocytes were: 0.1 (Advia 2120i) and 0.3 × 10 9 /L (XN‐1000, BC‐5310CRP, DxH 900). LoQ for platelets: 3 (Advia 2120i), 4 (XN‐1000), 5 (DxH 900), and 10 × 10 9 /L (BC‐5310CRP). Linearity was confirmed for all, wherein Mindray had slightly wider 95%‐CI. No carryover was observed.

Conclusion

While all analyzers demonstrated acceptable performance for the majority of evaluated specifications, Sysmex XN‐1000 and Beckman Coulter DxH 900 achieved the most favorable overall performance.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

HematologyHumansBlood Cell CountReproducibility of Results

Stichwörter der Autor:innen

Beckman Coulter DxH 900Mindray BC‐5310CRPSiemens Advia 2120iSysmex XN‐1000evaluationhematology analyzers

Autor:innen und Institutionen

  • Department of Medical Laboratory Diagnostics, University Hospital "Sveti Duh", Zagreb, Croatia.
  • Department of Medical Biochemistry and Hematology, University of Zagreb, Faculty of Pharmacy and Biochemistry, Zagreb, Croatia.
  • Department of Sport and Exercise Medicine, University of Zagreb, Faculty of Kinesiology, Zagreb, Croatia.

ORCID-Profile: Ana Nikler

Zitieren

APA

Nikler, A., Rajković, M. G., & Biljak, V. R. (2025). Comparative Evaluation of Four Hematology Analyzers: Siemens Advia 2120i, Sysmex XN ‐1000, Mindray BC ‐5310 CRP , and Beckman Coulter DxH 900. Do Similar Performances Equal Same Quality?. International Journal of Laboratory Hematology. https://doi.org/10.1111/ijlh.70025

Vancouver

Nikler A, Rajković MG, Biljak VR. Comparative Evaluation of Four Hematology Analyzers: Siemens Advia 2120i, Sysmex XN ‐1000, Mindray BC ‐5310 CRP , and Beckman Coulter DxH 900. Do Similar Performances Equal Same Quality?. International Journal of Laboratory Hematology. 2025. doi:10.1111/ijlh.70025

BibTeX
@article{Nikler2025,
  author = {Nikler, Ana and Rajković, Marija Grdić and Biljak, Vanja Radišić},
  title = {Comparative Evaluation of Four Hematology Analyzers: Siemens Advia 2120i, Sysmex XN ‐1000, Mindray BC ‐5310 CRP , and Beckman Coulter DxH 900. Do Similar Performances Equal Same Quality?},
  journal = {International Journal of Laboratory Hematology},
  year = {2025},
  doi = {10.1111/ijlh.70025}
}
Kostenlos starten →