Assessment of Surface Roughness and Early Bacterial Adhesion on Machined, SLA, Anodized, and Laser-Modified Titanium Implant Surfaces: An In-Vitro Study
Journal of Pioneering Medical Sciences · 15.04.2026 · 15 (S01) · 94-98
DOI: 10.47310/jpms202515s0119 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Pioneering Medical Sciences, online erschienen am 15.04.2026. Verfasst von Anas Abdul Khader.
Abstract (Originalsprache)
Background
Implant surface characteristics influence both osseointegration and early microbial colonization. The present in-vitro study compared machined, sandblasted and acid-etched (SLA), anodized, and laser-modified titanium surfaces to determine how different modification protocols affect roughness and 24-hour bacterial adhesion.
Methods
Eighty commercially pure titanium grade 4 discs were randomly allocated to four groups (n=20 each): machined, SLA, anodized, and laser-modified. Surface characterization was performed using optical profilometry (Ra, Rz, Rq, Rsk, and Rku) and scanning electron microscopy. Early bacterial adhesion at 24 hours was assessed using Streptococcus mutans under aerobic conditions and Porphyromonas gingivalis under anaerobic conditions. Colony-forming units (CFU/mm²) and fluorescence microscopy coverage percentages quantified adhesion. One-way ANOVA, Tukey post-hoc testing, and Pearson correlation analysis were applied at a significance level of p<0.05.
Results
SLA surfaces exhibited the highest roughness (Ra 2.34 ± 0.28 μm), followed by laser-modified (1.87 ± 0.23 μm), anodized (1.12 ± 0.18 μm), and machined surfaces (0.43 ± 0.09 μm) (p<0.001). SLA also showed the greatest bacterial adhesion. Compared with SLA, laser-modified surfaces demonstrated approximately 38.2% lower S. mutans counts and 38.8% lower P. gingivalis counts despite remaining moderately rough. Strong positive correlations were observed between Ra and bacterial adhesion parameters (r=0.80-0.85; p<0.001).
Conclusion
Surface modification protocols significantly influence both roughness and early bacterial adhesion on titanium. Within the limitations of this 24-hour single-species in-vitro model, laser-modified surfaces showed a favorable balance between moderate roughness and lower bacterial adhesion than SLA surfaces, and therefore appear promising rather than definitive for future implant surface design.
Zitieren
APA
Khader, A. A. (2026). Assessment of Surface Roughness and Early Bacterial Adhesion on Machined, SLA, Anodized, and Laser-Modified Titanium Implant Surfaces: An In-Vitro Study. Journal of Pioneering Medical Sciences, 15(S01), 94-98. https://doi.org/10.47310/jpms202515s0119
Vancouver
Khader AA. Assessment of Surface Roughness and Early Bacterial Adhesion on Machined, SLA, Anodized, and Laser-Modified Titanium Implant Surfaces: An In-Vitro Study. Journal of Pioneering Medical Sciences. 2026;15(S01):94-98. doi:10.47310/jpms202515s0119
BibTeX
@article{Khader2026,
author = {Khader, Anas Abdul},
title = {Assessment of Surface Roughness and Early Bacterial Adhesion on Machined, SLA, Anodized, and Laser-Modified Titanium Implant Surfaces: An In-Vitro Study},
journal = {Journal of Pioneering Medical Sciences},
year = {2026},
volume = {15},
number = {S01},
pages = {94--98},
doi = {10.47310/jpms202515s0119}
}