Performance of Unmodified Mechanical Ventilators With 2% Hydrogen Gas Mixtures
Respiratory Care · 2025-04 · 70 (4) · 377-383
DOI: 10.4187/respcare.12371 ↗Kurz gefasst
Fachartikel in der Zeitschrift Respiratory Care, online erschienen am 02.04.2025. Verfasst von Julia Garcia Mancebo und 7 weiteren Autor:innen. Beteiligte Einrichtung: Dr. Mancebo is affiliated with Department of Cardiology, Boston Children's Hospital, Boston, Massachusetts; Division of Critical Care, Texas Children…. Hauptthemen laut MEDLINE: Hydrogen, Ventilators, Mechanical, Respiration, Artificial. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NHLBI NIH HHS.
Abstract (Originalsprache)
Background
Molecular hydrogen (H 2 ) is a breathable gas that has been shown to have anti-oxidative, anti-inflammatory, and anti-apoptotic properties that may positively impact ischemia-reperfusion injury. The provision of 2% H 2 through unmodified mechanical ventilators may facilitate the clinical translation of H 2 as a therapeutic in critical illness. The effect of 2% H 2 on ventilator performance is unknown.
Methods
Unmodified Maquet Servo-i, Maquet Servo-u, Dräger Evita Infinity V500, and Dräger Evita Babylog VN500 ventilators from clinical stock were tested in an experimental closed system using certified, premixed air and O 2 containing 2% H 2 gas. Wall air and O 2 supply were used as control. Ventilator settings were varied across the spectrum of neonatal to adult settings. End points included (1) difference between set and delivered tidal volume (V T ) (Douglas method), (2) difference between set versus delivered O 2 concentration, (3) delivered H 2 concentration (gas chromatography), and (4) ventilator pre-use check malfunction. Correlation between set and measured end points were quantified by linear regression analysis and bias by Bland-Altman analysis.
Results
During H 2 administration, the average bias in measured versus set V T was within ± 10% for all ventilators except for the Babylog VN500, which exhibited an average bias of −89.2% (95% CI −107.0 to −71.3). The average bias in measured F IO 2 was within ± 10% of set for all ventilators. Except for the Babylog VN500, all ventilators passed the pre-use check.
Conclusions
Unmodified Servo-i, Servo-u, and Evita V500 ventilators deliver 2% H 2 mixtures with acceptable accuracy in V T and F IO 2 . The Babylog VN500, which uses hot-wire anometry and a higher set operating temperature, exhibits unacceptably inaccurate delivery of V T with H 2 mixtures.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Dr. Mancebo is affiliated with Department of Cardiology, Boston Children's Hospital, Boston, Massachusetts; Division of Critical Care, Texas Children's Hospital, Houston, Texas; and Department of Pediatrics, Baylor Coll…
- Ms. Sack and Dr. Peng are affiliated with Department of Cardiology, Boston Children's Hospital, Boston, Massachusetts.
- Mr. Nguyen is affiliated with Department of Respiratory Therapy, Texas Children's Hospital, Houston, Texas.
- Ms. Sosa and Drs. Anders and Roddy are affiliated with Division of Critical Care, Texas Children's Hospital, Houston, Texas; and Department of Pediatrics, Baylor College of Medicine, Houston, Texas.
- Dr. Kheir is affiliated with Department of Cardiology, Boston Children's Hospital, Boston, Massachusetts; and Department of Pediatrics, Harvard Medical School, Boston, Massachusetts.
Förderung
- NHLBI NIH HHS
Zitieren
APA
Mancebo, J. G., Sack, K., Nguyen, R., Peng, Y., Sosa, S., Anders, M., Roddy, D. J., & Kheir, J. N. (2025). Performance of Unmodified Mechanical Ventilators With 2% Hydrogen Gas Mixtures. Respiratory Care, 70(4), 377-383. https://doi.org/10.4187/respcare.12371
Vancouver
Mancebo JG, Sack K, Nguyen R, Peng Y, Sosa S, Anders M, et al. Performance of Unmodified Mechanical Ventilators With 2% Hydrogen Gas Mixtures. Respiratory Care. 2025;70(4):377-383. doi:10.4187/respcare.12371
BibTeX
@article{Mancebo2025,
author = {Mancebo, Julia Garcia and Sack, Kristen and Nguyen, Richard and Peng, Yifeng and Sosa, Syndy and Anders, Marc and Roddy, Dantin J and Kheir, John N},
title = {Performance of Unmodified Mechanical Ventilators With 2% Hydrogen Gas Mixtures},
journal = {Respiratory Care},
year = {2025},
volume = {70},
number = {4},
pages = {377--383},
doi = {10.4187/respcare.12371}
}