Improved efficiency of the new Mirus™ system challenged by ventilatory settings—a bench study

Frederic Albrecht, Claudia Schirra, Kathrin Scheffler, Thomas Volk, Andreas Meiser

Journal of Clinical Monitoring and Computing · 2025

DOI: 10.1007/s10877-025-01344-5 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Monitoring and Computing, 2025. Verfasst von Frederic Albrecht und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anesthesiology, Intensive Care and Pain Therapy, Saarland University Medical Center and Saarland University Faculty of Medicine, Hombur…. Hauptthemen laut MEDLINE: Anesthetics, Inhalation, Respiration, Artificial, Ventilators, Mechanical, Anesthesia, Inhalation. Die Arbeit nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Universitätsklinikum des Saarlandes und Medizinische Fakultät der Universität des Saarlandes.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The Mirus™ system (Technologie Institut Medizin GmbH, Germany) enables target-controlled delivery of volatile anaesthetics with open system ventilators. Its interface, positioned between Y-piece and the patient, injects anaesthetic vapor during inspiration, while adsorbing anaesthetic and resupplying it during the next inspiration. A newly introduced interface (Lisa-44) was evaluated under body temperature pressure saturated and normocapnic conditions using a test lung. Volume-controlled ventilation was applied with a 500 mL tidal volume, 10 bpm respiratory rate, and inspiratory flow (IF) times of 2.5 s and 0.5 s. Isoflurane (sevoflurane) target concentrations were stepwise increased to 1.5 MAC and then decreased. An external gas monitor recorded real-time anaesthetic concentrations. End-tidal concentrations (EC) and area under the curve (AUC) were analyzed for each breath. Accuracy and precision were assessed using Bland-Altman plots. Consumption was calculated and compared to historical controls. Mirus EC measurements correlated well with external gas monitor readings. ECs fluctuated around targets, with higher targets requiring more frequent injections and larger AUCs. Anaesthetic consumption was lower than historical controls and comparable to the Sedaconda ACD. At 0.3–0.5 MAC, hourly isoflurane consumption ranged from 1.0 to 1.7 mL (sevoflurane: 3.8–6.1 mL). However, very short IF times significantly increased consumption, reaching 43 mL sevoflurane at an EC of 2.6 vol%, with prolonged injections extending into the expiration phase. The new system demonstrates high accuracy, precision, and improved efficiency, suggesting reduced anaesthetic consumption in clinical use. However, high concentrations combined with very short IF times substantially increase consumption, indicating potential limitations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Anesthetics, InhalationRespiration, ArtificialVentilators, MechanicalAnesthesia, InhalationHumansIsofluraneSevofluraneEquipment DesignTidal VolumeMethyl EthersReproducibility of ResultsArea Under Curve

Stichwörter der Autor:innen

Inhaled sedationIsofluraneMirusSevofluraneVolatile anaesthetics

Substanzen

Anesthetics, InhalationIsofluraneSevofluraneMethyl Ethers

Autor:innen und Institutionen

  • Department of Anesthesiology, Intensive Care and Pain Therapy, Saarland University Medical Center and Saarland University Faculty of Medicine, Homburg, Germany. Frederic.Albrecht@uks.eu.
  • Department of Anesthesiology, Intensive Care and Pain Therapy, Saarland University Medical Center and Saarland University Faculty of Medicine, Homburg, Germany.
  • Outcomes Research Consortium®, Houston, TX, USA.

Förderung

  • Universitätsklinikum des Saarlandes und Medizinische Fakultät der Universität des Saarlandes

Zitieren

APA

Albrecht, F., Schirra, C., Scheffler, K., Volk, T., & Meiser, A. (2025). Improved efficiency of the new Mirus™ system challenged by ventilatory settings—a bench study. Journal of Clinical Monitoring and Computing. https://doi.org/10.1007/s10877-025-01344-5

Vancouver

Albrecht F, Schirra C, Scheffler K, Volk T, Meiser A. Improved efficiency of the new Mirus™ system challenged by ventilatory settings—a bench study. Journal of Clinical Monitoring and Computing. 2025. doi:10.1007/s10877-025-01344-5

BibTeX
@article{Albrecht2025,
  author = {Albrecht, Frederic and Schirra, Claudia and Scheffler, Kathrin and Volk, Thomas and Meiser, Andreas},
  title = {Improved efficiency of the new Mirus™ system challenged by ventilatory settings—a bench study},
  journal = {Journal of Clinical Monitoring and Computing},
  year = {2025},
  doi = {10.1007/s10877-025-01344-5}
}
Kostenlos starten →