Evaluation of a new hyperbaric oxygen ventilator during pressure-controlled ventilation
Diving and Hyperbaric Medicine Journal · 2024
DOI: 10.28920/dhm54.3.212-216 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Diving and Hyperbaric Medicine Journal, 2024. Verfasst von Cong Wang und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Hyperbaric Oxygen, Beijing Tiantan Hospital, Capital Medical University, A zone, No.199 Nansihuan West Road, Fengtai District, Beijing …. Hauptthemen laut MEDLINE: Hyperbaric Oxygenation, Ventilators, Mechanical, Tidal Volume. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Introduction
The stability of a new hyperbaric ventilator (Shangrila590, Beijing Aeonmed Company, Beijing, China) at different clinically relevant pressures in a hyperbaric chamber during pressure-controlled ventilation (PCV) was investigated.
Methods
The ventilator was connected to a test lung in the multiplace hyperbaric chamber. The inspiratory pressure (PI) of the ventilator was set to 1.0, 1.5, 2.0, 2.5 and 3.0 kPa (approximately 10, 15, 20, 25 and 30 cmH₂O). The compliance and resistance of the test lung were set to 200 mL·kPa⁻¹ and 2 kPa·L⁻¹·s⁻¹, respectively. Experiments were conducted at 101, 203 and 284 kPa ambient pressure (1.0, 2.0 and 2.8 atmospheres absolute respectively). At each of the 5 PI values, the tidal volume (VT), peak inspiratory pressure (Ppeak) and peak inspiratory flow (Fpeak) displayed by the ventilator and the test lung were recorded for 20 cycles. Test lung data were considered the actual ventilation values. The ventilation data were compared among the three groups to evaluate the stability of the ventilator.
Results
At every PI, the Ppeak detected by the ventilator decreased slightly with increasing ambient pressure. The Fpeak values measured by the test lung decreased substantially as the ambient pressure increased. Nevertheless, the reduction in VT at 284 kPa and PI 30 cmH₂O (compared to performance at 101 kPa) was comparatively small (approximately 60 ml).
Conclusions
In PCV mode this ventilator provided relatively stable VT across clinically relevant PI values to ambient pressures as high as 284 kPa. However, because Fpeak decreases at higher ambient pressure, some user adjustment might be necessary for precise VT maintenance during clinical use at higher PIs and ambient pressures.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Hyperbaric Oxygen, Beijing Tiantan Hospital, Capital Medical University, A zone, No.199 Nansihuan West Road, Fengtai District, Beijing 100070, China.
- Department of Hyperbaric Oxygen, Huazhong University of Science and Technology Union Shenzhen Hospital (Nanshan Hospital), No. 89 Taoyuan Road, Nanshan District, Shenzhen 518052, China.
- Corresponding author: Professor Lianbi Xue, Department of Hyperbaric Oxygen, Huazhong University of Science and Technology Union Shenzhen Hospital (Nanshan Hospital), No. 89 Taoyuan Road, Nanshan District, Shenzhen 5180…
Zitieren
APA
Wang, C., Yu, Q., Liu, Y., Ren, Z., Liu, Y., & Xue, L. (2024). Evaluation of a new hyperbaric oxygen ventilator during pressure-controlled ventilation. Diving and Hyperbaric Medicine Journal. https://doi.org/10.28920/dhm54.3.212-216
Vancouver
Wang C, Yu Q, Liu Y, Ren Z, Liu Y, Xue L. Evaluation of a new hyperbaric oxygen ventilator during pressure-controlled ventilation. Diving and Hyperbaric Medicine Journal. 2024. doi:10.28920/dhm54.3.212-216
BibTeX
@article{Wang2024,
author = {Wang, Cong and Yu, Qiuhong and Liu, Yaling and Ren, Ziqi and Liu, Ying and Xue, Lianbi},
title = {Evaluation of a new hyperbaric oxygen ventilator during pressure-controlled ventilation},
journal = {Diving and Hyperbaric Medicine Journal},
year = {2024},
doi = {10.28920/dhm54.3.212-216}
}
EUVIMED