Comparison of High-Frequency Oscillatory Ventilators

Kaoru Okazaki, Jumpei Kuroda

Respiratory Care · 2024

DOI: 10.4187/respcare.10773 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Respiratory Care, 2024. Verfasst von Kaoru Okazaki und 1 weiteren Autor:innen. Beteiligte Einrichtung: Drs Okazaki and Kuroda are affiliated with the Department of Neonatology, Tokyo Metropolitan Children's Medical Center, Tokyo, Japan. kaoru_okazaki@t…. Hauptthemen laut MEDLINE: High-Frequency Ventilation. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The performance of high-frequency oscillatory ventilators (HFOV) differs by the waveform generation mode and circuit characteristics. Few studies have described the performance of piston-type HFOV. The present study aimed to compare the amplitude required to reach the target high-frequency tidal volume ( V T ); determine the relationship between the settings and actual pressure in amplitude or mean airway pressure ( P ¯ aw ); and describe the interaction among compliance, frequency, and endotracheal tube (ETT) inner diameter in 4 HFOV models, including Humming X, Vue (a piston type ventilator commonly used in Japan), VN500 (a diaphragm type), and SLE5000 (a reverse jet type). METHODS: The oscillatory ventilators were evaluated by using a 50-mL test lung with 0.5 and 1.0 mL/cm H 2 O compliance, P ¯ aw of 10 cm H 2 O, frequency of 12 and 15 Hz, and ETT inner diameters 2.0, 2.5, and 3.5 mm. At each permutation of compliance, frequency, and ETT, the target high-frequency V T was increased from 0.5 to 3.0 mL. The change in P ¯ aw from the ventilator (ventilator P ¯ aw ) to Y-piece (Y P ¯ aw ) and alveolar pressure (alveolar P ¯ aw ) and the change in amplitude from the ventilator (ventilator amplitude) to Y-piece (Y amplitude) and alveolar pressure (alveolar amplitude) were determined at high-frequency V T of 1.0 and 3.0 mL. RESULTS: To achieve the target high-frequency V T , the Humming X and Vue required a higher amplitude than did the SLE5000, but the maximum amplitude in the VN500 was unable to attain a larger high-frequency V T . Ventilator P ¯ aw and alveolar pressure decreased at the Y-piece with the Humming X and Vue but increased with the SLE5000. The ventilator P ¯ aw in the VN500 decreased remarkably at a frequency of 15 Hz. The ventilator amplitude in all 4 ventilators decreased while temporarily increasing at the Y-piece in the VN500. CONCLUSIONS: The actual measured value, such as alveolar P ¯ aw and high-frequency V T , varied according to the type of HFOV system and the inner diameter of the ETT, even with identical settings. Clinicians should therefore determine the setting appropriate to each HFOV model.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

High-Frequency VentilationHumansLungVentilators, MechanicalTidal VolumePressure

Stichwörter der Autor:innen

amplitudehigh-frequency oscillatory ventilationhigh-frequency tidal volumemean airway pressureneonate

Autor:innen und Institutionen

  • Drs Okazaki and Kuroda are affiliated with the Department of Neonatology, Tokyo Metropolitan Children's Medical Center, Tokyo, Japan. kaoru_okazaki@tmhp.jp.
  • Drs Okazaki and Kuroda are affiliated with the Department of Neonatology, Tokyo Metropolitan Children's Medical Center, Tokyo, Japan.

Zitieren

APA

Okazaki, K., & Kuroda, J. (2024). Comparison of High-Frequency Oscillatory Ventilators. Respiratory Care. https://doi.org/10.4187/respcare.10773

Vancouver

Okazaki K, Kuroda J. Comparison of High-Frequency Oscillatory Ventilators. Respiratory Care. 2024. doi:10.4187/respcare.10773

BibTeX
@article{Okazaki2024,
  author = {Okazaki, Kaoru and Kuroda, Jumpei},
  title = {Comparison of High-Frequency Oscillatory Ventilators},
  journal = {Respiratory Care},
  year = {2024},
  doi = {10.4187/respcare.10773}
}
Kostenlos starten →