1-4-2: Evaluation of applied mechanical power to individual lungs in a simulator-based setting of one ventilator for two patients
PLOS One · 2025
DOI: 10.1371/journal.pone.0328813 ↗Kurz gefasst
Fachartikel in der Zeitschrift PLOS One, 2025. Verfasst von Lars-Olav Harnisch und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesiology, University Medical Center, Georg-August-University, Göttingen, Germany. Hauptthemen laut MEDLINE: Ventilators, Mechanical, Respiration, Artificial, Lung. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The concept of ventilating multiple patients concurrently using a single ventilator has been proposed as a solution when the demand for ventilators surpasses the available supply. While the practicality of this approach has been established, a thorough evaluation of the risks involved has yet to be comprehensively addressed.
Methods
Two circuits, a simple one (circuit-1) and another with an adjustable resistance valve (circuit-2), were evaluated within an experimental framework utilizing two computer-controlled lung simulators (TestChest and ASL 5000). These simulators were ventilated by an ICU (intensive care unit) ventilator (Servo-u) employing various ventilation modes (volume- and pressure-controlled ventilation). The study was conducted under differing respiratory conditions, characterized by low compliance (20 ml/cmH 2 O) as well as normal-high compliance (100 ml/cmH 2 O), in order to ascertain the applied tidal volume (VT), pressures, and the resultant mechanical power (MP).
Results
Circuit-1: The applied VT, pressures, and MP differed significantly between the two simulators, as well as in relation to ventilation mode, compliance, and respiratory rate (RR) (p < 0.001); the differences were most pronounced in settings with differing compliance levels. Circuit-2: Differences in VT, pressures, and MP were observed between simulators concerning valve settings (p < 0.001). The VT demonstrated a negative correlation, with volumes derived from valve closure spanning from 50 to 100 ml across all settings. In the design of circuit-2, MP exceeded the 12 J/min threshold in both lung simulators at elevated RR and could only be decreased through valve closure followed by a consequential hypoventilation in one simulator.
Conclusion
The simultaneous ventilation of two patients using a single ventilator is technically viable, yet it presents considerable risks. Even with the integration of an adjustable resistance valve to accommodate varying lung complexit i es, the likelihood of unilateral hypoventilation and elevated mechanical stress remains high.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Anaesthesiology, University Medical Center, Georg-August-University, Göttingen, Germany.
- Department of Anesthesiology, Perioperative and Pain Medicine, Icahn School of Medicine at Mount Sinai, New York, New York, United States of America.
- Department of Medical Statistics, University Medical Center, Georg-August-University Göttingen, Göttingen, Germany.
- Outcomes Research Consortium, Cleveland, Ohio, United States of America.
ORCID-Profile: Lars-Olav Harnisch, Leif Saager
Zitieren
APA
Harnisch, L. O., Czock, C., Levin, M. A., Wieditz, J., Saager, L., Meissner, K., & Moerer, O. (2025). 1-4-2: Evaluation of applied mechanical power to individual lungs in a simulator-based setting of one ventilator for two patients. PLOS One. https://doi.org/10.1371/journal.pone.0328813
Vancouver
Harnisch LO, Czock C, Levin MA, Wieditz J, Saager L, Meissner K, et al. 1-4-2: Evaluation of applied mechanical power to individual lungs in a simulator-based setting of one ventilator for two patients. PLOS One. 2025. doi:10.1371/journal.pone.0328813
BibTeX
@article{Harnisch2025,
author = {Harnisch, Lars-Olav and Czock, Christian and Levin, Matthew A and Wieditz, Johannes and Saager, Leif and Meissner, Konrad and Moerer, Onnen},
title = {1-4-2: Evaluation of applied mechanical power to individual lungs in a simulator-based setting of one ventilator for two patients},
journal = {PLOS One},
year = {2025},
doi = {10.1371/journal.pone.0328813}
}
EUVIMED