AbstractBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Non-tunneled hemodialysis catheters (HDCs) are commonly used to initiate renal replacement therapy in hospitalized patients, especially those with acute kidney injury. HDC malposition is associated with delayed therapy and inadequate dialysis. Agitated saline bubble-enhanced ultrasound (SBUS) has emerged as a method for visualizing catheter position at the time of insertion. This study aimed to analyze the accuracy and concordance of SBUS and chest radiography (CXR) for assessing HDC positioning according to the internal jugular vein (IJV) insertion site. A total of 152 hospitalized patients undergoing IJV catheter insertion were assessed from December 2019 to May 2021. Time from puncture to CXR result, procedure-related complications, blood flow during dialysis, and the accuracy and concordance of SBUS compared with CXR for determining HDC position adequacy were evaluated. The median time between catheter insertion and CXR was 191 min [interquartile range (IQR), 93-288 min], with no difference between insertion sites. The main complication after catheter insertion was arterial puncture (2% overall), with a similar incidence at both sites. Inadequate catheter blood flow was more frequent with left IJV insertion than with right IJV insertion. For right IJV catheters, concordance between SBUS and CXR was high (Gwet's AC1 = 0.95; 95% confidence interval [CI], 0.915-0.992), with an accuracy of 95.6%. For left IJV catheters, concordance was low (Gwet's AC1 = 0.37; 95% CI, -0.21-0.94), with an accuracy of 64.3%. Compared with CXR, SBUS was suitable for assessing catheter placement following right IJV insertion but may be less suitable following left IJV insertion.

Cite

APA

Barbosa, G. S. B., Ribeiro, R. G., Cortes, D. D. P. V. R., Filho, M. P. M., Botter, D. A., Gutierrez, G. M., Sandoval, M. C., Smolentzov, I., & Rodrigues, C. E. (2026). Accuracy and Concordance of Agitated Saline Bubble-Enhanced Ultrasound versus Chest Radiography by Internal Jugular Insertion Site. Journal of Visualized Experiments. https://doi.org/10.3791/72476

Vancouver

Barbosa GSB, Ribeiro RG, Cortes DDPVR, Filho MPM, Botter DA, Gutierrez GM, et al. Accuracy and Concordance of Agitated Saline Bubble-Enhanced Ultrasound versus Chest Radiography by Internal Jugular Insertion Site. Journal of Visualized Experiments. 2026. doi:10.3791/72476

BibTeX
@article{Barbosa2026,
  author = {Barbosa, Géssica Sabrine Braga and Ribeiro, Rayra Gomes and Cortes, Daniela del Pilar Via Reque and Filho, Marcelo Paes Menezes and Botter, Denise Aparecida and Gutierrez, Gabriel Monteiro and Sandoval, Monica Carneiro and Smolentzov, Igor and Rodrigues, Camila Eleutério},
  title = {Accuracy and Concordance of Agitated Saline Bubble-Enhanced Ultrasound versus Chest Radiography by Internal Jugular Insertion Site},
  journal = {Journal of Visualized Experiments},
  year = {2026},
  number = {237},
  doi = {10.3791/72476}
}
Start for free →