Compact and Real-Time Radiation Dosimeter Using Silicon Photomultipliers for In Vivo Dosimetry in Radiation Therapy
Sensors · 2025
DOI: 10.3390/s25030857 ↗Kurz gefasst
Fachartikel in der Zeitschrift Sensors, 2025. Verfasst von Jeongho Kim und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Samsung Changwon Hospital, Sungkyunkwan University School of Medicine, Changwon 51353, Republic of Korea. Hauptthemen laut MEDLINE: Silicon, Radiation Dosimeters, In Vivo Dosimetry, Radiometry. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea (NRF) and funded by the Korean Government (MIST), Korea Institute of Energy Technology and Planning(KETEP) grant funded by the Korean gov-ernment (MOTIE, National Research Foundation of Korea (NRF).
Abstract (Originalsprache)
Existing dosimeters for radiation therapy are typically large, and their performance in in vivo system applications has not been assessed. This study develops a compact real-time dosimeter using silicon photomultipliers, plastic scintillators, and optical fibers and evaluates its in vivo applicability for radiation therapy. Dose calibration, dose-rate dependency and linearity, and short-term repeatability tests were conducted using solid water phantoms and bolus materials, and in vivo dosimetry was performed using an in-house phantom. The characterization evaluation results showed high linearity, with a coefficient of determination of 0.9995 for dose rates of 100–600 monitoring units (MU)/min, confirming an error rate within 2% when converted to dosage. In the short-term repeatability tests, the dosimeter exhibited good characteristics, with relative standard deviation (RSD) values lower than 2% for each beam delivery and an RSD value of 0.03% over ten beam deliveries. Dose measurements using the phantom indicated an average error rate of 3.83% compared to the values calculated using the treatment planning system. These results demonstrate a performance comparable to that of commercial metal-oxide-semiconductor field-effect transistors and plastic scintillator-based dosimeters. Based on these findings, the developed dosimeter has significant potential for in vivo radiation therapy applications.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiation Oncology, Samsung Changwon Hospital, Sungkyunkwan University School of Medicine, Changwon 51353, Republic of Korea.
- Marine Radioactivity Monitoring Group, Korea Marine Environment Management Co., Busan 48931, Republic of Korea.
- Interdisciplinary Program in Precision Public Health, Korea University, Seoul 02841, Republic of Korea.
- Department of Chemistry, Konkuk University, Seoul 05029, Republic of Korea.
- Department of Radiation Oncology, Jeju National University Hospital, Jeju 63241, Republic of Korea.
ORCID-Profile: Jeongho Kim, Beomjun Park
Förderung
- National Research Foundation of Korea (NRF) and funded by the Korean Government (MIST)
- Korea Institute of Energy Technology and Planning(KETEP) grant funded by the Korean gov-ernment (MOTIE
- National Research Foundation of Korea (NRF)
- Korean Government (MIST)
- Korea Institute of Energy Technology and Planning (KETEP)
Zitieren
APA
Kim, J., Park, J., Park, B., Kim, Y., Park, B., & Park, S. H. (2025). Compact and Real-Time Radiation Dosimeter Using Silicon Photomultipliers for In Vivo Dosimetry in Radiation Therapy. Sensors. https://doi.org/10.3390/s25030857
Vancouver
Kim J, Park J, Park B, Kim Y, Park B, Park SH. Compact and Real-Time Radiation Dosimeter Using Silicon Photomultipliers for In Vivo Dosimetry in Radiation Therapy. Sensors. 2025. doi:10.3390/s25030857
BibTeX
@article{Kim2025,
author = {Kim, Jeongho and Park, Jeehoon and Park, Byungdo and Kim, Yonghoon and Park, Beomjun and Park, So Hyun},
title = {Compact and Real-Time Radiation Dosimeter Using Silicon Photomultipliers for In Vivo Dosimetry in Radiation Therapy},
journal = {Sensors},
year = {2025},
doi = {10.3390/s25030857}
}
EUVIMED