Characterization of a flexible a‐Si:H detector for in vivo dosimetry in therapeutic x‐ray beams
Medical Physics · 2024
DOI: 10.1002/mp.17013 ↗Kurz gefasst
Fachartikel in der Zeitschrift Medical Physics, 2024. Verfasst von Matthew James Large und 46 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia. Hauptthemen laut MEDLINE: Silicon, Radiometry. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The increasing use of complex and high dose‐rate treatments in radiation therapy necessitates advanced detectors to provide accurate dosimetry. Rather than relying on pre‐treatment quality assurance (QA) measurements alone, many countries are now mandating the use of in vivo dosimetry, whereby a dosimeter is placed on the surface of the patient during treatment. Ideally, in vivo detectors should be flexible to conform to a patient's irregular surfaces.
Purpose
This study aims to characterize a novel hydrogenated amorphous silicon (a‐Si:H) radiation detector for the dosimetry of therapeutic x‐ray beams. The detectors are flexible as they are fabricated directly on a flexible polyimide (Kapton) substrate.
Methods
The potential of this technology for application as a real‐time flexible detector is investigated through a combined dosimetric and flexibility study. Measurements of fundamental dosimetric quantities were obtained including output factor (OF), dose rate dependence (DPP), energy dependence, percentage depth dose (PDD), and angular dependence. The response of the a‐Si:H detectors investigated in this study are benchmarked directly against commercially available ionization chambers and solid‐state diodes currently employed for QA practices.
Results
The a‐Si:H detectors exhibit remarkable dose linearities in the direct detection of kV and MV therapeutic x‐rays, with calibrated sensitivities ranging from (0.580 ± 0.002) pC/cGy to (19.36 ± 0.10) pC/cGy as a function of detector thickness, area, and applied bias. Regarding dosimetry, the a‐Si:H detectors accurately obtained OF measurements that parallel commercially available detector solutions. The PDD response closely matched the expected profile as predicted via Geant4 simulations, a PTW Farmer ionization chamber and a PTW ROOS chamber. The most significant variation in the PDD performance was 5.67%, observed at a depth of 3 mm for detectors operated unbiased. With an external bias, the discrepancy in PDD response from reference data was confined to ± 2.92% for all depths (surface to 250 mm) in water‐equivalent plastic. Very little angular dependence is displayed between irradiations at angles of 0° and 180°, with the most significant variation being a 7.71% decrease in collected charge at a 110° relative angle of incidence. Energy dependence and dose per pulse dependence are also reported, with results in agreement with the literature. Most notably, the flexibility of a‐Si:H detectors was quantified for sample bending up to a radius of curvature of 7.98 mm, where the recorded photosensitivity degraded by (−4.9 ± 0.6)% of the initial device response when flat. It is essential to mention that this small bending radius is unlikely durin g in vivo patient dosimetry. In a more realistic scenario, with a bending radius of 15–20 mm, the variation in detector response remained within ± 4%. After substantial bending, the detector's photosensitivity when returned to a flat condition was (99.1 ± 0.5)% of the original response.
Conclusions
This work successfully characterizes a flexible detector based on thin‐film a‐Si:H deposited on a Kapton substrate for applications in therapeutic x‐ray dosimetry. The detectors exhibit dosimetric performances that parallel commercially available dosimeters, while also demonstrating excellent flexibility results.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Centre for Medical Radiation Physics, University of Wollongong, Wollongong, New South Wales, Australia.
- School of Physics, Najran University, Najran, Saudi Arabia.
- Shoalhaven Cancer Care Centre, Nowra, New South Wales, Australia.
- Ecole Polytechnique Fédérale de Lausanne (EPFL), Photovoltaics and Thin-Film Electronics Laboratory (PV-Lab), Neuchâtel, Switzerland.
- INFN Sezione di Lecce, via per Arnesano, Lecce, Italy.
- Department of Mathematics and Physics "Ennio de Giorgi", University of Salento, Via per Arnesano, Lecce, Italy.
Zitieren
APA
Large, M. J., Bashiri, A., Dookie, Y., McNamara, J., Antognini, L., Aziz, S., Calcagnile, L., Caricato, A. P., Catalano, R., Chila, D., Cirrone, G. A. P., Croci, T., Cuttone, G., Dunand, S., Fabi, M., Frontini, L., Grimani, C., Ionica, M., Kanxheri, K., …, & Petasecca, M. (2024). Characterization of a flexible a‐Si:H detector for in vivo dosimetry in therapeutic x‐ray beams. Medical Physics. https://doi.org/10.1002/mp.17013
Vancouver
Large MJ, Bashiri A, Dookie Y, McNamara J, Antognini L, Aziz S, et al. Characterization of a flexible a‐Si:H detector for in vivo dosimetry in therapeutic x‐ray beams. Medical Physics. 2024. doi:10.1002/mp.17013
BibTeX
@article{Large2024,
author = {Large, Matthew James and Bashiri, Aishah and Dookie, Yashiv and McNamara, Joanne and Antognini, Luca and Aziz, Saba and Calcagnile, Lucio and Caricato, Anna Paola and Catalano, Roberto and Chila, Deborah and Cirrone, Giuseppe Antonio Pablo and Croci, Tomasso and Cuttone, Giacomo and Dunand, Sylvain and Fabi, Michele and Frontini, Luca and Grimani, Catia and Ionica, Maria and Kanxheri, Keida and Liberali, Valentino and Maurizio, Martino and Maruccio, Giuseppe and Mazza, Giovanni and Menichelli, Mauro and Monteduro, Anna Grazia and Morozzi, Arianna and Moscatelli, Francesco and Pallotta, Stefania and Passeri, Daniele and Pedio, Maddalena and Petringa, Giada and Peverini, Francesca and Piccolo, Lorenzo and Placidi, Pisana and Quarta, Gianluca and Rizzato, Silvia and Sabbatini, Federico and Servoli, Leonello and Stabile, Alberto and Talamonti, Cinzia and Thomet, Jonathan Emanuel and Tosti, Luca and Villani, Mattia and Wheadon, Richard James and Wyrsch, Nicolas and Zema, Nicola and Petasecca, Marco},
title = {Characterization of a flexible a‐Si:H detector for in vivo dosimetry in therapeutic x‐ray beams},
journal = {Medical Physics},
year = {2024},
doi = {10.1002/mp.17013}
}
EUVIMED