A scintillation dosimeter with real‐time positional tracking information for in vivo dosimetry error detection in HDR brachytherapy

Daline Tho iD, Marie-Claude Lavallée, Luc Beaulieu

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.14150 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Daline Tho und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: In Vivo Dosimetry, Brachytherapy. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Sciences and Engineering Research Council of Canada (NSERC).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate the performance of an electromagnetic (EM)‐tracked scintillation dosimeter in detecting source positional errors of IVD in HDR brachytherapy treatment. Materials and Methods Two different scintillator dosimeter prototypes were coupled to 5 degrees‐of‐freedom (DOF) EM sensors read by an Aurora V3 system. The scintillators used were a 0.3 × 0.4 × 0.4 mm 3 ZnSe:O and a BCF‐60 plastic scintillator of 0.5 mm diameter and 2.0 mm in length (Saint‐Gobain Crystals). The sensors were placed at the dosimeter's tip at 20.0 mm from the scintillator. The EM sampling rate was 40/s while the scintillator signal was sampled at 100 000/s using two photomultiplier tubes from Hamamatsu (series H10722) connected to a data acquisition board. A high‐pass filter and a low‐pass filter were used to separate the light signal into two different channels. All measurements were performed with an afterloader unit (Flexitron‐Elekta AB, Sweden) in full‐scattered (TG43) conditions. EM tracking was further used to provide distance/angle‐dependent energy correction for the ZnSe:O inorganic scintillator. For the error detection part, lateral shifts of 0.5 to 3 mm were induced by moving the source away from its planned position. Indexer length (longitudinal) errors between 0.5 to 10 mm were also introduced. The measured dose rate difference was converted to a shift distance, with and without using the positional information from the EM sensor.

Results

The inorganic scintillator had both a signal‐to‐noise‐ratio (SNR) and signal‐to‐background‐ratio (SBR) close to 70 times higher than those of the plastic scintillator. The mean absolute difference from the dose measurement to the dose calculated with TG‐43U1 was 1.5% ±0.7%. The mean absolute error for BCF‐60 detector was 1.7% when compared to TG‐43 calculations formalism. With the inorganic scintillator and EM tracking, a maximum area under the curve (AUC) gain of 24.0% was obtained for a 0.5‐mm lateral shift when using the EMT data with the ZnSe:O. Lower AUC gains were obtained for a 3‐mm lateral shifts with both scintillators. For the plastic scintillator, the highest gain from using EM tracking information occurred for a 0.5‐mm lateral shift at 20 mm from the source. The maximal gain (17.4%) for longitudinal errors was found at the smallest shifts (0.5 mm).

Conclusions

This work demonstrates that integrating EM tracking to in vivo scintillation dosimeters enables the detection of smaller shifts, by decreasing the dosimeter positioning uncertainty. It also serves to perform position‐dependent energy correction for the inorganic scintillator,providing better SNR and SBR, allowing detection of errors at greater distances from the source.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

In Vivo DosimetryBrachytherapyHumansScintillation CountingRadiation DosimetersElectromagnetic PhenomenaRadiometryRadiotherapy Dosage

Stichwörter der Autor:innen

HDRbrachytherapyelectromagnetic trackingplastic scintillation dosimeter

Autor:innen und Institutionen

  • Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas, USA.
  • Département de physique, de génie physique et d'optique, et Centre de recherche sur le cancer, Université Laval, Québec, Québec, Canada.
  • Service de physique médicale et de radioprotection, Centre intégré de cancérologie, CHU de Québec-Université Laval et Centre de recherche du CHU de Québec, Québec, Canada.

ORCID-Profile: Daline Tho

Förderung

  • National Sciences and Engineering Research Council of Canada (NSERC)

Zitieren

APA

Tho, D., Lavallée, M. C., & Beaulieu, L. (2023). A scintillation dosimeter with real‐time positional tracking information for in vivo dosimetry error detection in HDR brachytherapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14150

Vancouver

Tho D, Lavallée MC, Beaulieu L. A scintillation dosimeter with real‐time positional tracking information for in vivo dosimetry error detection in HDR brachytherapy. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14150

BibTeX
@article{Tho2023,
  author = {Tho, Daline and Lavallée, Marie-Claude and Beaulieu, Luc},
  title = {A scintillation dosimeter with real‐time positional tracking information for in vivo dosimetry error detection in HDR brachytherapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.14150}
}
Kostenlos starten →