EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Remote Estimation of Blood Pressure Using Millimeter-Wave Frequency-Modulated Continuous-Wave Radar

Lovedeep Singh, Sungjin You iD, Byung Jang Jeong, Chiwan Koo iD, Youngwook Kim iD

Sensors · 2023

DOI: 10.3390/s23146517 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Sensors, 2023. Verfasst von Lovedeep Singh und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Electrical and Computer Engineering, California State University, Fresno, CA 93740, USA. Hauptthemen laut MEDLINE: Radar, Pulse Wave Analysis. Die Arbeit wurde bisher 29-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Electronics and Telecommunications Research Institute, National Research Foundation (NRF).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper proposes to remotely estimate a human subject’s blood pressure using a millimeter-wave radar system. High blood pressure is a critical health threat that can lead to diseases including heart attacks, strokes, kidney disease, and vision loss. The commonest method of measuring blood pressure is based on a cuff that is contact-based, non-continuous, and cumbersome to wear. Continuous remote monitoring of blood pressure can facilitate early detection and treatment of heart disease. This paper investigates the possibility of using millimeter-wave frequency-modulated continuous-wave radar to measure the heart blood pressure by means of pulse wave velocity (PWV). PWV is known to be highly correlated with blood pressure, which can be measured by pulse transit time. We measured PWV using a two-millimeter wave radar focused on the subject’s chest and wrist. The measured time delay provided the PWV given the length from the chest to the wrist. In addition, we analyzed the measured radar signal from the wrist because the shape of the pulse wave purveyed information on blood pressure. We investigated the area under the curve (AUC) as a feature and found that AUC is strongly correlated with blood pressure. In the experiment, five human subjects were measured 50 times each after performing different activities intended to influence blood pressure. We used artificial neural networks to estimate systolic blood pressure (SBP) and diastolic blood pressure (SBP) with both PWV and AUC as inputs. The resulting root mean square errors of estimated blood pressure were 3.33 mmHg for SBP and 3.14 mmHg for DBP.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadarPulse Wave AnalysisHumansBlood PressureVital SignsBlood Pressure Determination

Stichwörter der Autor:innen

FMCW radarblood pressurepulse pressurepulse transit timepulse wave velocity

Autor:innen und Institutionen

  • Department of Electrical and Computer Engineering, California State University, Fresno, CA 93740, USA.
  • Electronics and Telecommunications Research Institute, Daejeon 34129, Republic of Korea.
  • Department of Electronic Engineering, Hanbat National University, Daejeon 34158, Republic of Korea.
  • Department of Electronic Engineering, Sogang University, Seoul 04107, Republic of Korea.

ORCID-Profile: Sungjin You, Chiwan Koo, Youngwook Kim

Förderung

  • Electronics and Telecommunications Research Institute
  • National Research Foundation (NRF)

Zitieren

APA

Singh, L., You, S., Jeong, B. J., Koo, C., & Kim, Y. (2023). Remote Estimation of Blood Pressure Using Millimeter-Wave Frequency-Modulated Continuous-Wave Radar. Sensors. https://doi.org/10.3390/s23146517

Vancouver

Singh L, You S, Jeong BJ, Koo C, Kim Y. Remote Estimation of Blood Pressure Using Millimeter-Wave Frequency-Modulated Continuous-Wave Radar. Sensors. 2023. doi:10.3390/s23146517

BibTeX
@article{Singh2023,
  author = {Singh, Lovedeep and You, Sungjin and Jeong, Byung Jang and Koo, Chiwan and Kim, Youngwook},
  title = {Remote Estimation of Blood Pressure Using Millimeter-Wave Frequency-Modulated Continuous-Wave Radar},
  journal = {Sensors},
  year = {2023},
  doi = {10.3390/s23146517}
}
Kostenlos starten →