Distinct Structural Determinants of Failure in Morse Taper and Internal Hex Implant Systems
Journal of Functional Biomaterials · 17.07.2026 · 17 (7) · 346
DOI: 10.3390/jfb17070346 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Functional Biomaterials, online erschienen am 17.07.2026. Verfasst von Sergio Alexandre Gehrke und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pharmaceutical Science, School of Health Sciences, Universidade do Vale do Itajai, Itajai 88302-901, Brazil. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
To evaluate the influence of implant–abutment connection design, implant diameter, and simulated marginal bone loss on fracture resistance and failure patterns of dental implant systems. Materials and Methods: A total of 180 implant–abutment assemblies were tested, including Morse taper (MT) and internal hex (IH) connections with diameters of 3.5, 4.0, and 5.0 mm. Implants were embedded at two simulated bone levels (0 and 3 mm) and loaded at 30° until failure, in accordance with ISO 14801:2015. Fracture resistance (N) was analyzed using three-way ANOVA.
Results
Connection type, implant diameter, bone level, and their interactions significantly affected fracture resistance (p < 0.001). Internal hex implants showed a marked diameter-dependent increase in resistance and greater reduction under simulated bone loss, particularly in reduced diameters. In contrast, Morse taper implants demonstrated similar resistance values regardless of implant diameter or bone level. Failure patterns also differed between systems: internal hex implants exhibited cervical implant fractures in reduced diameters, whereas Morse taper implants showed progressive abutment deformation and/or abutment fracture without implant body fracture.
Conclusions
Implant fracture resistance is strongly influenced by implant–abutment connection geometry. Within the limitations of this static in vitro study, MT systems demonstrated diameter-independent mechanical stability and prosthetic-controlled failure patterns, whereas IH systems were highly sensitive to diameter reduction and simulated bone loss.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Pharmaceutical Science, School of Health Sciences, Universidade do Vale do Itajai, Itajai 88302-901, Brazil.
- Department of Implantology, Bioface/Postgrados en Odontologia/Universidad Catolica de Murcia, Montevideo 11100, Uruguay.
- Department of Implantology, Be Dental School, Universidade do Vale do Itajai (UNIVALI), Saint Joseph 88102-700, Brazil.
- Department of Innovative Technologies in Medicine & Dentistry, University of Chieti-Pescara, 66013 Chieti, Italy.
- GF10 Foundation, St. Louis, MO 63104, USA.
- Periodontics, Bitonte College of Dentistry, Northeast Ohio Medical University (NEOMED), Rootstown, OH 44272, USA.
ORCID-Profile: Márcio de Carvalho Formiga, Antonio Scarano, Juliana Campos Hasse Fernandes, Gustavo Vicentis Oliveira Fernandes, Fátima de Campos Buzzi
Zitieren
APA
Gehrke, S. A., Coura, G., Mello, B. F., de Carvalho Formiga, M., Scarano, A., Fernandes, J. C. H., Fernandes, G. V. O., & de Campos Buzzi, F. (2026). Distinct Structural Determinants of Failure in Morse Taper and Internal Hex Implant Systems. Journal of Functional Biomaterials, 17(7), 346. https://doi.org/10.3390/jfb17070346
Vancouver
Gehrke SA, Coura G, Mello BF, de Carvalho Formiga M, Scarano A, Fernandes JCH, et al. Distinct Structural Determinants of Failure in Morse Taper and Internal Hex Implant Systems. Journal of Functional Biomaterials. 2026;17(7):346. doi:10.3390/jfb17070346
BibTeX
@article{Gehrke2026,
author = {Gehrke, Sergio Alexandre and Coura, Gustavo and Mello, Bruno Freitas and de Carvalho Formiga, Márcio and Scarano, Antonio and Fernandes, Juliana Campos Hasse and Fernandes, Gustavo Vicentis Oliveira and de Campos Buzzi, Fátima},
title = {Distinct Structural Determinants of Failure in Morse Taper and Internal Hex Implant Systems},
journal = {Journal of Functional Biomaterials},
year = {2026},
volume = {17},
number = {7},
pages = {346},
doi = {10.3390/jfb17070346}
}
EUVIMED