Influence of Implant Macro-Design, -Length, and -Diameter on Primary Implant Stability Depending on Different Bone Qualities Using Standard Drilling Protocols—An In Vitro Analysis

Milan Stoilov iD, Ramin Shafaghi, Helmut Stark, Michael Marder iD, Dominik Kraus, Norbert Enkling

Journal of Functional Biomaterials · 12.09.2023 · 14 (9) · 469

DOI: 10.3390/jfb14090469 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Functional Biomaterials, online erschienen am 12.09.2023. Verfasst von Milan Stoilov und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Prosthodontics, Preclinical Education and Dental Materials Science, Bonn University, 53111 Bonn, Germany. Die Arbeit wurde bisher 23-mal zitiert (Crossref) und nennt 80 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

(1)

Background

Primary implant stability is vital for successful implant therapy. This study explores the influence of implant shape, length, and diameter on primary stability in different bone qualities. (2)

Methods

Three implant systems (two parallel-walled and one tapered) with various lengths and diameters were inserted into polyurethane foam blocks of different densities (35, 25, 15, and 10 PCF) using standard drilling protocols. Primary stability was assessed through insertion torque (IT) and resonance frequency analysis (RFA). Optimal ranges were defined for IT (25 to 50 Ncm) and RFA (ISQ 60 to 80). A comparison of implant groups was conducted to determine adherence to the optimal ranges. (3)

Results

Implant macro-design, -length, and -diameter and bone block density significantly influenced IT and RFA. Optimal IT was observed in 8/40 and 9/40 groups for the parallel-walled implants, while the tapered implant achieved optimal IT in 13/40 groups (within a 25–50 Ncm range). Implant diameter strongly impacted primary stability, with sufficient stability achieved in only one-third of cases despite the tapered implant’s superiority. (4)

Conclusions

The findings highlight the need to adapt the drilling protocol based on diverse bone qualities in clinical practice. Further investigations should explore the impact of these adapted protocols on implant outcomes.

Schlagwörter

Stichwörter der Autor:innen

ISQRFAbone qualityimmediate loadinginsertion torqueprimary stability

Autor:innen und Institutionen

  • Department of Prosthodontics, Preclinical Education and Dental Materials Science, Bonn University, 53111 Bonn, Germany.
  • Department of Reconstructive Dentistry and Gerodontology, Bern University, 3012 Bern, Switzerland.

ORCID-Profile: Milan Stoilov, Michael Marder

Zitieren

APA

Stoilov, M., Shafaghi, R., Stark, H., Marder, M., Kraus, D., & Enkling, N. (2023). Influence of Implant Macro-Design, -Length, and -Diameter on Primary Implant Stability Depending on Different Bone Qualities Using Standard Drilling Protocols—An In Vitro Analysis. Journal of Functional Biomaterials, 14(9), 469. https://doi.org/10.3390/jfb14090469

Vancouver

Stoilov M, Shafaghi R, Stark H, Marder M, Kraus D, Enkling N. Influence of Implant Macro-Design, -Length, and -Diameter on Primary Implant Stability Depending on Different Bone Qualities Using Standard Drilling Protocols—An In Vitro Analysis. Journal of Functional Biomaterials. 2023;14(9):469. doi:10.3390/jfb14090469

BibTeX
@article{Stoilov2023,
  author = {Stoilov, Milan and Shafaghi, Ramin and Stark, Helmut and Marder, Michael and Kraus, Dominik and Enkling, Norbert},
  title = {Influence of Implant Macro-Design, -Length, and -Diameter on Primary Implant Stability Depending on Different Bone Qualities Using Standard Drilling Protocols—An In Vitro Analysis},
  journal = {Journal of Functional Biomaterials},
  year = {2023},
  volume = {14},
  number = {9},
  pages = {469},
  doi = {10.3390/jfb14090469}
}
Kostenlos starten →