Evaluation of the Effects of Dental Implant Dimension and Design on Primary Implant Stability Using Different Stability Measurement Devices

Tuna Sumer iD, Nihat Akbulut iD

Journal of Clinical Medicine · 05.11.2026 · 15 (21) · 8626

DOI: 10.3390/jcm15218626 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 05.11.2026. Verfasst von Tuna Sumer und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Institute of Graduate Studies, Ondokuz Mayis University, Kurupelit, Samsun 55139,…. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Ondokuz Mayıs University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background/Objectives: Dental implants are an established treatment for functional and aesthetic rehabilitation of partially or completely edentulous patients. This in vitro study evaluates the effects of dental implant design, diameter, length, and bone density on primary implant stability using Osstell ISQ and Penguin Resonance Frequency Analysis (RFA) devices. Materials and Methods: A total of 96 dental implants were included in this study. Combinations of two implant designs (T6 Standard and T6 Torq), two implant diameters (3.5 mm and 5.0/4.8 mm), and two implant lengths (8 mm and 14 mm) were placed into synthetic bone blocks with densities of PCF15 and PCF30. Implant stability was measured twice from two different directions using Osstell ISQ and Penguin RFA devices. The data were statistically analyzed.

Results

Both RFA devices demonstrated high measurement reproducibility. When all measurements were evaluated collectively, the intraclass correlation coefficient (ICC) reached 0.999 for both devices. Bone density was identified as the most influential factor affecting primary implant stability. Significantly higher ISQ values were obtained in the high-density bone model (p < 0.001). Increases in implant length and diameter significantly enhanced primary stability (p < 0.001). T6 Torq implants exhibited significantly higher ISQ values than T6 Standard implants in low-density bone and small-dimension implant combinations (p < 0.001).

Conclusions

Under the present experimental conditions, bone density showed the greatest influence on primary implant stability, followed by implant length and implant diameter. In low-density bone, particularly with short and narrow implants, more aggressive implant macrodesigns may contribute to higher primary stability. However, the clinical relevance of these findings requires further investigation.

Schlagwörter

Stichwörter der Autor:innen

bone densitydental implantsprimary stabilityresonance frequency analysis

Autor:innen und Institutionen

  • Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Institute of Graduate Studies, Ondokuz Mayis University, Kurupelit, Samsun 55139, Turkey.
  • Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Ankara University, Mevlana Blv. Yenimahalle, Ankara 06560, Turkey.

ORCID-Profile: Tuna Sumer, Nihat Akbulut

Förderung

  • Ondokuz Mayıs University

Zitieren

APA

Sumer, T., & Akbulut, N. (2026). Evaluation of the Effects of Dental Implant Dimension and Design on Primary Implant Stability Using Different Stability Measurement Devices. Journal of Clinical Medicine, 15(21), 8626. https://doi.org/10.3390/jcm15218626

Vancouver

Sumer T, Akbulut N. Evaluation of the Effects of Dental Implant Dimension and Design on Primary Implant Stability Using Different Stability Measurement Devices. Journal of Clinical Medicine. 2026;15(21):8626. doi:10.3390/jcm15218626

BibTeX
@article{Sumer2026,
  author = {Sumer, Tuna and Akbulut, Nihat},
  title = {Evaluation of the Effects of Dental Implant Dimension and Design on Primary Implant Stability Using Different Stability Measurement Devices},
  journal = {Journal of Clinical Medicine},
  year = {2026},
  volume = {15},
  number = {21},
  pages = {8626},
  doi = {10.3390/jcm15218626}
}
Kostenlos starten →