Effects of Two Dental Implant Micromotor Systems for Dental Implant Placement on Implant Stability and Removal Torque: An Animal Experiment

Keunbada Son iD, Young-Tak Son iD, Sung-Min Hwang iD, Jae Mok Lee, Jin-Wook Kim, Kyu-Bok Lee iD

Materials · 29.08.2025 · 18 (17) · 4048

DOI: 10.3390/ma18174048 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Materials, online erschienen am 29.08.2025. Verfasst von Keunbada Son und 5 weiteren Autor:innen. Beteiligte Einrichtung: Advanced Dental Device Development Institute (A3DI), Kyungpook National University, Daegu 41940, Republic of Korea. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 57 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Research Foundation of Korea, Korea Evaluation Institute of Industrial Technology, Korean government.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This in vivo animal study aimed to evaluate the effects of two different implant placement micromotor systems on implant stability and removal torque. In a within-animal crossover design, twenty titanium implants (AnyOne fixture; internal type; diameter, 3.5 mm; length, 7.0 mm; Megagen, Daegu, Republic of Korea) were placed in the tibiae of five rabbits using a conventional micromotor system (NSK group: SurgicPro+; NSK, Kanuma, Japan) and a diode laser-integrated micromotor system (SAESHIN group: BLP 10; Saeshin, Daegu, Republic of Korea). Resonance frequency analysis provided the implant stability quotient (ISQ) immediately after placement and at four weeks. Micro-computed tomography quantified the bone–implant interface gap (BIG). Removal torque was measured at sacrifice. Linear mixed-effects models with a random intercept for rabbit generated adjusted means with 95% confidence intervals (CIs) (α = 0.05). Equivalence for the four-week ISQ used two one-sided tests with a margin of ±5 ISQ. The SAESHIN group achieved a higher immediate ISQ than the NSK group (difference =+6.9 ISQ; 95% CI +1.3–+12.5; p = 0.018). At four weeks, the ISQ did not differ (difference = −1.2 ISQ; 95% CI −4.3–+1.9; p = 0.42), and equivalence was supported (TOST p_lower = 0.024; p_upper = 0.019). Removal torque was comparable (difference = +4.3 N·cm; 95% CI −5.2–+13.8; p = 0.36). BIG metrics showed no between-system differences across regions. ICC indicated clustering for ISQ and torque (0.36 and 0.31). The diode laser-integrated micromotor system yielded a higher immediate ISQ under a standardized 35 N·cm seating torque, whereas the ISQ, removal torque, and BIG at four weeks were comparable to those of the conventional system. The immediate ISQ should be interpreted as stiffness under fixed torque rather than superior device-dependent interlocking. These findings support the clinical interchangeability of the two systems for early osseointegration endpoints in preclinical settings.

Schlagwörter

Stichwörter der Autor:innen

bone–implant interface gapdental implant micromotorimplant stability quotientmicro-computed tomographyremoval torqueresonance frequency analysis

Autor:innen und Institutionen

  • Advanced Dental Device Development Institute (A3DI), Kyungpook National University, Daegu 41940, Republic of Korea.
  • Department of Dental Science, Graduate School, Kyungpook National University, Daegu 41940, Republic of Korea.
  • Department of Periodontology, School of Dentistry, Kyungpook National University, Daegu 41566, Republic of Korea.
  • Department of Oral & Maxillofacial Surgery, School of Dentistry, Kyungpook National University, Daegu 41566, Republic of Korea.
  • Department of Prosthodontics, School of Dentistry, Kyungpook National University, Daegu 41940, Republic of Korea.

ORCID-Profile: Keunbada Son, Young-Tak Son, Sung-Min Hwang, Kyu-Bok Lee

Förderung

  • National Research Foundation of Korea
  • Korea Evaluation Institute of Industrial Technology
  • Korean government
  • Ministry of Trade, Industry and Energy

Zitieren

APA

Son, K., Son, Y. T., Hwang, S. M., Lee, J. M., Kim, J. W., & Lee, K. B. (2025). Effects of Two Dental Implant Micromotor Systems for Dental Implant Placement on Implant Stability and Removal Torque: An Animal Experiment. Materials, 18(17), 4048. https://doi.org/10.3390/ma18174048

Vancouver

Son K, Son YT, Hwang SM, Lee JM, Kim JW, Lee KB. Effects of Two Dental Implant Micromotor Systems for Dental Implant Placement on Implant Stability and Removal Torque: An Animal Experiment. Materials. 2025;18(17):4048. doi:10.3390/ma18174048

BibTeX
@article{Son2025,
  author = {Son, Keunbada and Son, Young-Tak and Hwang, Sung-Min and Lee, Jae Mok and Kim, Jin-Wook and Lee, Kyu-Bok},
  title = {Effects of Two Dental Implant Micromotor Systems for Dental Implant Placement on Implant Stability and Removal Torque: An Animal Experiment},
  journal = {Materials},
  year = {2025},
  volume = {18},
  number = {17},
  pages = {4048},
  doi = {10.3390/ma18174048}
}
Kostenlos starten →