Evaluating dental implant stability using three devices Osstell ® , Periotest ® , and AnyCheck ® : a clinical study
Journal of Oral Medicine and Oral Surgery · 2024 · 30 (3) · 20
DOI: 10.1051/mbcb/2024023 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Oral Medicine and Oral Surgery, online erschienen am 16.10.2024. Verfasst von Alamin Y. Dhahi und 1 weiteren Autor:innen. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Implant stability is usually measured with resonance frequency analysis (RFA) and damping capacity assessment (DCA). This study aimed to measure primary and secondary stabilities using 3 devices that are based on these methods, namely; RFA (Osstell ® ) and DCA (Periotest ® and AnyCheck ® ) to assess the correlations of the measurements obtained by these devices and the correlations between implant stability and insertion torque. Material and Methods: This observational prospective study included 35 dental implants. The implant stability was measured using the 3 devices. Mann–Whitney U test and unpaired t -test assessed the relationship between implant stability and insertion torque, while the Spearman and Pearson correlations measured the correlation between readings collected via the 3 devices for the primary and secondary stabilities.
Results
For the primary stability, there was a strong positive correlation between Osstell ® and AnyCheck ® and moderate negative correlations between Periotest ® and both Osstell ® and AnyCheck ® . While for the secondary stability, strong correlations with similar patterns were observed among the 3 devices. The stability measurements showed significant relationships with the insertion torque.
Conclusions
The 3 devices are reliable in measuring implant stability; also, high insertion torque can lead to improved implant stabilities (primary and secondary).
Zitieren
APA
Dhahi, A. Y., & Bede, S. Y. (2024). Evaluating dental implant stability using three devices Osstell ® , Periotest ® , and AnyCheck ® : a clinical study. Journal of Oral Medicine and Oral Surgery, 30(3), 20. https://doi.org/10.1051/mbcb/2024023
Vancouver
Dhahi AY, Bede SY. Evaluating dental implant stability using three devices Osstell ® , Periotest ® , and AnyCheck ® : a clinical study. Journal of Oral Medicine and Oral Surgery. 2024;30(3):20. doi:10.1051/mbcb/2024023
BibTeX
@article{Dhahi2024,
author = {Dhahi, Alamin Y. and Bede, Salwan Y.},
title = {Evaluating dental implant stability using three devices Osstell ® , Periotest ® , and AnyCheck ® : a clinical study},
journal = {Journal of Oral Medicine and Oral Surgery},
year = {2024},
volume = {30},
number = {3},
pages = {20},
doi = {10.1051/mbcb/2024023}
}