Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice
Journal of Clinical Medicine · 01.02.2024 · 13 (3) · 847
DOI: 10.3390/jcm13030847 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 01.02.2024. Verfasst von Michele Cennamo und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Translational Medical Sciences, University of Naples "Federico II", 80126 Naples, Italy. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
The erythrocyte sedimentation rate (ESR) is a routine and aspecific test that is still widely used. The reference-manual method for ESR determination is the Westergren method. The VES-MATIC 5 is a novel, fully automated, and closed system based on a modified Westergren method. This study conceived the aim of comparing two ESR analytical analysers, Test 1 and the VES-MATIC 5, with the reference method in routine practice.
Methods
This study included 264 randomly analysed samples. A comparison between the two methods and Westergren was performed, and they were evaluated for inter-run and intra-run precision. In addition, we investigated possible interferences and different sensitivities to conventional analytes.
Results
The comparison of methods by Passing–Bablok analysis provided a good agreement for both systems, with a better correlation for VES-MATIC 5 (p = 0.96) than Test 1 (p = 0.93), and sensitivity studies did not show any significant influence.
Conclusions
The VES-MATIC 5 analyser demonstrated excellent comparability with the reference method, and it had better performance than Test 1. It can be employed in routine practice, bringing advantages such as a reduction in the probability of human error compared to the manual method, as well as an increase in operator safety and environmental protection.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Translational Medical Sciences, University of Naples "Federico II", 80126 Naples, Italy.
- Clinical Pathology and Microbiology Unit, Laboratory Analysis, ASST Lariana, Hospital Sant'Anna, 22100 Como, Italy.
- Department of Diagnostic Innovation, Fondazione IRCCS Istituto Nazionale dei Tumori, 20133 Milan, Italy.
- Department of Diagnostics and Public Health, Section of Clinical Biochemistry, University of Verona, 37134 Verona, Italy.
- Department of Nephrology and Dialysis, ASST Lariana, Hospital Sant'Anna, 22100 Como, Italy.
ORCID-Profile: Fabio Bertani
Zitieren
APA
Cennamo, M., Giuliano, L., Arrigoni, G., Fardone, V., Russo, R., De Tomasi, L. M., Bertani, F., Cammarota, G., Brunetti, G., Del Vecchio, L., & Partenope, M. (2024). Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice. Journal of Clinical Medicine, 13(3), 847. https://doi.org/10.3390/jcm13030847
Vancouver
Cennamo M, Giuliano L, Arrigoni G, Fardone V, Russo R, De Tomasi LM, et al. Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice. Journal of Clinical Medicine. 2024;13(3):847. doi:10.3390/jcm13030847
BibTeX
@article{Cennamo2024,
author = {Cennamo, Michele and Giuliano, Loredana and Arrigoni, Gloria and Fardone, Valentina and Russo, Roberta and De Tomasi, Luca Maria and Bertani, Fabio and Cammarota, Gaetano and Brunetti, Giovanni and Del Vecchio, Lucia and Partenope, Michelarcangelo},
title = {Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice},
journal = {Journal of Clinical Medicine},
year = {2024},
volume = {13},
number = {3},
pages = {847},
doi = {10.3390/jcm13030847}
}