EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice

Michele Cennamo, Loredana Giuliano, Gloria Arrigoni, Valentina Fardone, Roberta Russo, Luca Maria De Tomasi, Fabio Bertani iD, Gaetano Cammarota, Giovanni Brunetti, Lucia Del Vecchio, Michelarcangelo Partenope

Journal of Clinical Medicine · 01.02.2024 · 13 (3) · 847

DOI: 10.3390/jcm13030847 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Clinical Medicine, online erschienen am 01.02.2024. Verfasst von Michele Cennamo und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Translational Medical Sciences, University of Naples "Federico II", 80126 Naples, Italy. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 59 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The erythrocyte sedimentation rate (ESR) is a routine and aspecific test that is still widely used. The reference-manual method for ESR determination is the Westergren method. The VES-MATIC 5 is a novel, fully automated, and closed system based on a modified Westergren method. This study conceived the aim of comparing two ESR analytical analysers, Test 1 and the VES-MATIC 5, with the reference method in routine practice.

Methods

This study included 264 randomly analysed samples. A comparison between the two methods and Westergren was performed, and they were evaluated for inter-run and intra-run precision. In addition, we investigated possible interferences and different sensitivities to conventional analytes.

Results

The comparison of methods by Passing–Bablok analysis provided a good agreement for both systems, with a better correlation for VES-MATIC 5 (p = 0.96) than Test 1 (p = 0.93), and sensitivity studies did not show any significant influence.

Conclusions

The VES-MATIC 5 analyser demonstrated excellent comparability with the reference method, and it had better performance than Test 1. It can be employed in routine practice, bringing advantages such as a reduction in the probability of human error compared to the manual method, as well as an increase in operator safety and environmental protection.

Schlagwörter

Stichwörter der Autor:innen

ICSHTest 1VES-MATIC 5Westergren methoderythrocyte sedimentation rate (ESR)

Autor:innen und Institutionen

  • Department of Translational Medical Sciences, University of Naples "Federico II", 80126 Naples, Italy.
  • Clinical Pathology and Microbiology Unit, Laboratory Analysis, ASST Lariana, Hospital Sant'Anna, 22100 Como, Italy.
  • Department of Diagnostic Innovation, Fondazione IRCCS Istituto Nazionale dei Tumori, 20133 Milan, Italy.
  • Department of Diagnostics and Public Health, Section of Clinical Biochemistry, University of Verona, 37134 Verona, Italy.
  • Department of Nephrology and Dialysis, ASST Lariana, Hospital Sant'Anna, 22100 Como, Italy.

ORCID-Profile: Fabio Bertani

Zitieren

APA

Cennamo, M., Giuliano, L., Arrigoni, G., Fardone, V., Russo, R., De Tomasi, L. M., Bertani, F., Cammarota, G., Brunetti, G., Del Vecchio, L., & Partenope, M. (2024). Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice. Journal of Clinical Medicine, 13(3), 847. https://doi.org/10.3390/jcm13030847

Vancouver

Cennamo M, Giuliano L, Arrigoni G, Fardone V, Russo R, De Tomasi LM, et al. Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice. Journal of Clinical Medicine. 2024;13(3):847. doi:10.3390/jcm13030847

BibTeX
@article{Cennamo2024,
  author = {Cennamo, Michele and Giuliano, Loredana and Arrigoni, Gloria and Fardone, Valentina and Russo, Roberta and De Tomasi, Luca Maria and Bertani, Fabio and Cammarota, Gaetano and Brunetti, Giovanni and Del Vecchio, Lucia and Partenope, Michelarcangelo},
  title = {Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice},
  journal = {Journal of Clinical Medicine},
  year = {2024},
  volume = {13},
  number = {3},
  pages = {847},
  doi = {10.3390/jcm13030847}
}