EUVIMEDEUVIMED entdecken
Crossref

Erythrocyte Sedimentation Rate measured by Automated Analysers and Manual Westergren’s Method – A Comparative Study

Karthika V., Shyamala Gowri M, D. Arunthamizhpraba, Soundarrajan T., K. Raghul

International Journal of Current Pharmaceutical Review and Research · 30.04.2026 · 18 (04)

DOI: 10.25258/ijcpr.18.4.158 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Current Pharmaceutical Review and Research, online erschienen am 30.04.2026. Verfasst von Karthika V. und 4 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Erythrocyte Sedimentation Rate (ESR) determined by Westergren’s method is used in diagnosis and monitoring of inflammatory activities; however, it has many limitations which include inherent and technical factors. Alternate methods have been introduced to overcome the limitations of the manual method. These new methods must be properly evaluated before introducing in clinical laboratories. Materials and Methods: A total of 436 randomly collected blood samples were assayed simultaneously by the standard Westergren’s method and two automated methods using Roller 20LC and Celltac α+ MEK-1305.

Results

Results of these assays were subjected to statistical analysis using a Spearman rank coefficient of correlation, the Bland-Altman statistical methods and Passing-Bablok regression method. The present study on comparison of manual Westergren’s with Roller 20LC and Celltac α+ MEK-1305 revealed that the Spearman rank correlation coefficient ‘ρ’ of 0.781 (95% confidence interval [CI] 0.742 to 0.815, P

Zitieren

APA

V., K., M, S. G., Arunthamizhpraba, D., T., S., & Raghul, K. (2026). Erythrocyte Sedimentation Rate measured by Automated Analysers and Manual Westergren’s Method – A Comparative Study. International Journal of Current Pharmaceutical Review and Research, 18(04). https://doi.org/10.25258/ijcpr.18.4.158

Vancouver

V. K, M SG, Arunthamizhpraba D, T. S, Raghul K. Erythrocyte Sedimentation Rate measured by Automated Analysers and Manual Westergren’s Method – A Comparative Study. International Journal of Current Pharmaceutical Review and Research. 2026;18(04). doi:10.25258/ijcpr.18.4.158

BibTeX
@article{V2026,
  author = {V., Karthika and M, Shyamala Gowri and Arunthamizhpraba, D. and T., Soundarrajan and Raghul, K.},
  title = {Erythrocyte Sedimentation Rate measured by Automated Analysers and Manual Westergren’s Method – A Comparative Study},
  journal = {International Journal of Current Pharmaceutical Review and Research},
  year = {2026},
  volume = {18},
  number = {04},
  doi = {10.25258/ijcpr.18.4.158}
}
Kostenlos starten →