Performance Evaluation of the Alifax Test1 2.0 for Erythrocyte Sedimentation Rate Testing
Clinical Laboratory · 2026 · 72 (11/2026)
DOI: 10.7754/clin.lab.2025.251101 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Clinical Laboratory, 2026. Verfasst von E Schapkaitz und 2 weiteren Autor:innen. Hauptthemen laut MEDLINE: Blood Sedimentation, Photometry.
Abstract (Originalsprache)
BACKGROUND: The erythrocyte sedimentation rate (ESR) is widely measured using modified or alternative methods. The upgraded Alifax TEST1 2.0 (Alifax, Polverara, Italy), based on capillary photometry, is designed to improve operator safety and reduce analysis time. This study evaluated its diagnostic performance. METHODS: Precision, comparison, carryover, stability and reference ranges were evaluated in accordance with the International Council for Standardization in Hematology. For comparison with the earlier reference TEST1 1.0, 90 samples from adult and pediatric patients were analyzed in parallel at three ESR measurement ranges (2 - 20, 21 - 60 and 61 - 120 mm/hour) in the analyzer's internal and external modes. RESULTS: The TEST1 2.0 demonstrated acceptable precision with Alifax latex quality control material and patient samples in the ESR range > 20 mm/hour. The mean ESR was 35.1 ± 32.1 mm/hour versus 35.3 ± 32.4 mm/hour for the reference method (p = 0.872), with a mean weighted percentage difference of -0.1 ± 7.6 mm/hour. In the 21 - 58 mm/hour range, the y-intercept was 20.3 (95% CI, 14.7 - 26.0) and the slope 0.5 (95% CI, 0.3 - 0.6), indicating both constant and proportional bias, with a mean difference of -1.8 ± 12.8 mm/hour. ESR measurements on the TEST1 2.0 remained stable for up to 24 hours at room temperature. CONCLUSIONS: In conclusion, the TEST1 2.0 provided rapid and precise ESR measurements, with overall perfor-mance comparable to the TEST1 1.0. Differences observed in the clinically significant range highlight the need for careful evaluation and continued performance monitoring in routine laboratory use.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Zitieren
APA
Schapkaitz, E., Baiden, A., & Raburabu, S. (2026). Performance Evaluation of the Alifax Test1 2.0 for Erythrocyte Sedimentation Rate Testing. Clinical Laboratory, 72(11/2026). https://doi.org/10.7754/clin.lab.2025.251101
Vancouver
Schapkaitz E, Baiden A, Raburabu S. Performance Evaluation of the Alifax Test1 2.0 for Erythrocyte Sedimentation Rate Testing. Clinical Laboratory. 2026;72(11/2026). doi:10.7754/clin.lab.2025.251101
BibTeX
@article{Schapkaitz2026,
author = {Schapkaitz, E and Baiden, A and Raburabu, S},
title = {Performance Evaluation of the Alifax Test1 2.0 for Erythrocyte Sedimentation Rate Testing},
journal = {Clinical Laboratory},
year = {2026},
volume = {72},
number = {11/2026},
doi = {10.7754/clin.lab.2025.251101}
}
EUVIMED