Clinical implications of lateral head mis-centering on eye lens dose in pediatric computed tomography examinations: a phantom study
Frontiers in Medicine · 30.07.2026 · 13
DOI: 10.3389/fmed.2026.1838211 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Medicine, online erschienen am 30.07.2026. Verfasst von Mohammad Aljamal und 8 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Allied Medical Sciences, Department of Medical Imaging, Arab American University, Jenin, Palestine. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background and aim Patient mis-centering during pediatric CT examinations can significantly affect radiation dose distribution, particularly to radiosensitive organs such as the eye lens. This study quantitatively evaluates the effect of lateral head mis-centering on eye lens dose and investigates discrepancies between measured and system-reported dose metrics using a low-cost pediatric head phantom. Materials and methods A low-cost tissue-equivalent pediatric head phantom was constructed locally and validated using CT number comparisons with reference tissue values. Eye lens dose was measured using a pencil ionization chamber at the isocenter and at ±1–3 cm lateral displacements on two CT systems. Measured dose length product (DLP) values were compared with CT-reported DLP and air kerma-derived estimates. The data was analyzed statistically using Pearson correlation, paired sample t-tests, and one-way ANOVA.
Results
Strong correlations were observed between isocenter and off-center doses ( r = 0.994–0.998, p < 0.001). Although positional differences were not statistically significant ( p > 0.05), a consistent asymmetric trend was observed, with higher doses on the right side. System-reported DLP substantially overestimated measured values (e.g., 720 vs. 304.83 mGy·cm).
Conclusion
Lateral mis-centering introduces systematic dose distribution asymmetry and significant discrepancies in CT dose reporting. These findings highlight the importance of accurate patient positioning and caution in interpreting system-reported dose metrics in pediatric imaging.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Allied Medical Sciences, Department of Medical Imaging, Arab American University, Jenin, Palestine.
- Department of Radiological Sciences, College of Health and Rehabilitation Sciences, Princess Nourah bint Abdulrahman University, Riyadh, Saudi Arabia.
- Department of Diagnostic Radiology, College of Applied Medical Sciences, Taibah University, Madinah, Saudi Arabia.
- Shahin X-ray and MRI Medical Center, Hebron, Palestine.
- Department of Internal Medicine, College of Medicine, Princess Nourah bint Abdulrahman University, Riyadh, Saudi Arabia.
- Department of Radiological Sciences, College of Applied Medical Sciences, King Khalid University, Abha, Saudi Arabia.
Zitieren
APA
Aljamal, M., Hamd, Z. Y., Gareeballah, A., Shaheen, M., Almudaimeegh, L., Ahmed, A. M., Alharbi, M., Abuthabet, A., & Daraghmeh, Z. (2026). Clinical implications of lateral head mis-centering on eye lens dose in pediatric computed tomography examinations: a phantom study. Frontiers in Medicine, 13. https://doi.org/10.3389/fmed.2026.1838211
Vancouver
Aljamal M, Hamd ZY, Gareeballah A, Shaheen M, Almudaimeegh L, Ahmed AM, et al. Clinical implications of lateral head mis-centering on eye lens dose in pediatric computed tomography examinations: a phantom study. Frontiers in Medicine. 2026;13. doi:10.3389/fmed.2026.1838211
BibTeX
@article{Aljamal2026,
author = {Aljamal, Mohammad and Hamd, Zuhal Y and Gareeballah, Awadia and Shaheen, Mousa and Almudaimeegh, Lama and Ahmed, Amna M and Alharbi, Mohammed and Abuthabet, Abbas and Daraghmeh, Zaid},
title = {Clinical implications of lateral head mis-centering on eye lens dose in pediatric computed tomography examinations: a phantom study},
journal = {Frontiers in Medicine},
year = {2026},
volume = {13},
doi = {10.3389/fmed.2026.1838211}
}
EUVIMED