Fachartikel in der Zeitschrift Radiation Protection Dosimetry, online erschienen am 03.11.2026. Verfasst von Natasa Danilovic und 6 weiteren Autor:innen. Beteiligte Einrichtung: University of Novi Sad, Faculty of Sciences, Department of Physics, Trg Dositeja Obradovica 4, 21000 Novi Sad, Serbia. Die Arbeit nennt 30 Literaturangaben. Gefördert von: Ministry of Science, Technological Development and Innovation of the Republic of Serbia.
Abstract (Originalsprache)
Personal dosemeters calibration relies on the air kerma conversion coefficients (CCs), which depends on the phantom and used exposure geometry. This study aims to calculate Hp(3) CCs for the computerized tomography (CT) phantom (16 cm diameter, 15 cm height) and compare them with CCs for head phantom (20 × 20 cm) given in the Part 3 of international standard ISO 4037 (ISO 4037-3) to improve eye lens dosimetry methodologies in Serbia. Monte Carlo radiation transport simulations using MCNP Version 6.1.0 were performed for S-Cs (662 keV) and S-Co (1.25 MeV) reference photon fields. Absorbed dose and air kerma values were derived to calculate Hp(3) CCs for both CT and ISO phantoms across all specified angles for two radiation qualities. Calculated CCs were compared against the CCs ISO phantom values and previously published data to evaluate the relative deviations. This work provides the necessary foundation to determine whether correction coefficients should be introduced during the calibration and type-testing of passive eye dosimetry systems. Highlights MCNP 6.1.0 was used to derive conversion coefficients for a computerized tomography (CT) head phantom.The CT phantom caused backscatter deficits in S-Cs and S-Co photon fields.Posterior irradiation led to up to 20% overestimation for S-Cs.Dimension-specific corrections are needed for angular response testing.
Autor:innen und Institutionen
- University of Novi Sad, Faculty of Sciences, Department of Physics, Trg Dositeja Obradovica 4, 21000 Novi Sad, Serbia.
- Department of Radiation and Environmental Protection, Vinca Institute of Nuclear Sciences, National Institute of the Republic of Serbia, University of Belgrade, P.O. box 522, 11001 Belgrade, Serbia.
ORCID-Profile: Natasa Danilovic
Förderung
- Ministry of Science, Technological Development and Innovation of the Republic of Serbia
Zitieren
APA
Danilovic, N., Stankovic Petrovic, J., Bozovic, P., Topalovic, D., Pepic, I., Todorovic, N., & Nikolov, J. (2026). Calculation of the conversion coefficient from kerma to eye lens dose for computerized tomography head phantom. Radiation Protection Dosimetry. https://doi.org/10.1093/rpd/ncag123
Vancouver
Danilovic N, Stankovic Petrovic J, Bozovic P, Topalovic D, Pepic I, Todorovic N, et al. Calculation of the conversion coefficient from kerma to eye lens dose for computerized tomography head phantom. Radiation Protection Dosimetry. 2026. doi:10.1093/rpd/ncag123
BibTeX
@article{Danilovic2026,
author = {Danilovic, Natasa and Stankovic Petrovic, Jelena and Bozovic, Predrag and Topalovic, Dusan and Pepic, Ivana and Todorovic, Natasa and Nikolov, Jovana},
title = {Calculation of the conversion coefficient from kerma to eye lens dose for computerized tomography head phantom},
journal = {Radiation Protection Dosimetry},
year = {2026},
doi = {10.1093/rpd/ncag123}
}
EUVIMED