Effect of printing layer thickness on the mechanical properties of a direct-printed orthodontic aligner resin: An in vitro study
APOS Trends in Orthodontics · 28.07.2026 · 16 · 252-259
DOI: 10.25259/apos_350_2025 ↗Kurz gefasst
Fachartikel in der Zeitschrift APOS Trends in Orthodontics, online erschienen am 28.07.2026. Verfasst von Kiriti Sai Reddy Karri und 5 weiteren Autor:innen. Die Arbeit nennt 18 Literaturangaben.
Abstract (Originalsprache)
Objectives
The Objective of the study is to examine how print layer settings—50 and 100 µm affect the mechanical behavior of a clinically used direct-printed aligner resin. Material and Methods: This in vitro study investigated how printing layer thickness influences the mechanical properties of a directly printed aligner resin (Graphy Tera Harz TA-28). A total of 60 specimens were fabricated using a Uniz Nbee printer and divided into two groups based on layer thickness (50 µm and 100 µm; n = 30 per group). Standardized specimens for tensile, flexural, and compressive testing (n = 10 per test per group) were designed from STL templates and printed under identical conditions to ensure consistency. All samples underwent uniform post-processing, including centrifugation, ultrasonic cleaning, and nitrogen-assisted UV curing. Mechanical testing was carried out using a universal testing machine in accordance with ISO standards: ISO 527-2 for tensile testing, ISO 178 for flexural testing, and ISO 604 for compressive testing. The primary outcomes were strength and modulus values for each test. Data were analyzed using independent t-tests or Mann–Whitney U tests depending on data normality, with statistical significance set at α = 0.05.
Results
Compressive strength was significantly higher for the 50 µm group (2.42 ± 0.61 MPa) compared with the 100 µm group (1.84 ± 0.37 MPa; p = 0.034). Conversely, the 100 µm specimens demonstrated a higher tensile modulus (114.12 ± 44.01 MPa) than the 50 µm specimens (72.14 ± 25.8 MPa; p = 0.01).
Conclusion
Layer thickness influences the mechanical behavior of the Graphy TA-28 resin. Thinner layers improved compressive performance, while thicker layers increased tensile stiffness. These findings suggest that print settings can be tailored to achieve desired aligner strength and flexibility.
Zitieren
APA
Karri, K. S. R., Manojna, N. L., C.V, P. P., Alla, R. K., Uyyala, V. L., & Haribabu, N. (2026). Effect of printing layer thickness on the mechanical properties of a direct-printed orthodontic aligner resin: An in vitro study. APOS Trends in Orthodontics, 16, 252-259. https://doi.org/10.25259/apos_350_2025
Vancouver
Karri KSR, Manojna NL, C.V PP, Alla RK, Uyyala VL, Haribabu N. Effect of printing layer thickness on the mechanical properties of a direct-printed orthodontic aligner resin: An in vitro study. APOS Trends in Orthodontics. 2026;16:252-259. doi:10.25259/apos_350_2025
BibTeX
@article{Karri2026,
author = {Karri, Kiriti Sai Reddy and Manojna, Nachu Lakshmi and C.V, Padma Priya and Alla, Rama Krishna and Uyyala, Vijaya Lakshmi and Haribabu, Nagisetti},
title = {Effect of printing layer thickness on the mechanical properties of a direct-printed orthodontic aligner resin: An in vitro study},
journal = {APOS Trends in Orthodontics},
year = {2026},
volume = {16},
pages = {252--259},
doi = {10.25259/apos_350_2025}
}
EUVIMED