Influence of material and tightening torque on the subsidence of implant scan bodies
Journal of Prosthodontic Research · 2026 · 70 (1) · 37-42
DOI: 10.2186/jpr.jpr_d_24_00224 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Prosthodontic Research, 2026. Verfasst von Daiki Morita und 5 weiteren Autor:innen. Beteiligte Einrichtung: Section of Fixed Prosthodontics, Division of Oral Rehabilitation, Faculty of Dental Science, Kyushu University, Fukuoka, Japan. Hauptthemen laut MEDLINE: Torque, Dental Abutments, Dental Implants, Dental Implant-Abutment Design. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 33 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: To compare abutment and scan body subsidence and determine the appropriate tightening torque value for a scan body that can accurately replicate abutment subsidence. METHODS: An internally tapered connection implant was placed in a bone model. The titanium abutment (Ti Abut), polyetheretherketone scan body (PEEK SB), and titanium scan body (Ti SB) were tightened to torque values of 10 or 35 Ncm, with the 10 Ncm models used as reference models. Both the 10 and 35 Ncm models were scanned using a laboratory scanner. The vertical displacements of the Ti Abut, PEEK SB, and Ti SB at each torque value were determined using three-dimensional analysis software. RESULTS: Upon tightening the scan bodies and abutment to 35 Ncm, the displacement was significantly greater than when tightened to 10 Ncm (P < 0.01). At 35 Ncm, Ti SB and Ti Abut showed significantly greater displacements than PEEK SB, whereas no significant difference was observed between Ti SB and Ti Abut. CONCLUSIONS: The titanium scan bodies, when tightened at 35 Ncm, accurately replicated titanium abutment subsidence, unlike the PEEK scan bodies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Section of Fixed Prosthodontics, Division of Oral Rehabilitation, Faculty of Dental Science, Kyushu University, Fukuoka, Japan.
- Mechanical and Control Engineering, Kyushu Institute of Technology, Kitakyushu, Japan.
- Section of Implant and Rehabilitative Dentistry, Division of Oral Rehabilitation, Faculty of Dental Science, Kyushu University, Fukuoka, Japan.
Zitieren
APA
Morita, D., Matsuzaki, T., Sakai, N., Ogino, Y., Atsuta, I., & Ayukawa, Y. (2026). Influence of material and tightening torque on the subsidence of implant scan bodies. Journal of Prosthodontic Research, 70(1), 37-42. https://doi.org/10.2186/jpr.jpr_d_24_00224
Vancouver
Morita D, Matsuzaki T, Sakai N, Ogino Y, Atsuta I, Ayukawa Y. Influence of material and tightening torque on the subsidence of implant scan bodies. Journal of Prosthodontic Research. 2026;70(1):37-42. doi:10.2186/jpr.jpr_d_24_00224
BibTeX
@article{Morita2026,
author = {Morita, Daiki and Matsuzaki, Tatsuya and Sakai, Nobuo and Ogino, Yoichiro and Atsuta, Ikiru and Ayukawa, Yasunori},
title = {Influence of material and tightening torque on the subsidence of implant scan bodies},
journal = {Journal of Prosthodontic Research},
year = {2026},
volume = {70},
number = {1},
pages = {37--42},
doi = {10.2186/jpr.jpr_d_24_00224}
}
EUVIMED