Effect of material type, torque value, and sterilization on linear displacements of a scan body: An in vitro study
Clinical Implant Dentistry and Related Research · 2023
DOI: 10.1111/cid.13187 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Implant Dentistry and Related Research, 2023. Verfasst von Emre Diker und 5 weiteren Autor:innen. Beteiligte Einrichtung: Private Practice, Med Dis Dental Clinic, Ankara, Turkey. Hauptthemen laut MEDLINE: Dental Implants. Die Arbeit wurde bisher 28-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
There is limited knowledge on the effect of scan body (SB) material type, torque value, and sterilization on linear displacements of implant SBs.
Purpose
To evaluate the effect of material type, torque value, and sterilization on linear displacements of SBs during screw tightening by using digital image correlation (DIC) analysis. Materials and Methods One polyetheretherketone (PEEK, Zfx Intraoral Scan Body) and one titanium SB (Ti, MPS Zimmer Scanbody R1410) were tightened with 5 Ncm torque on two implants (Zimmer TSV ⌀4.7 mm) by using a digital torque limiting device. SBs' initial spatial positions relative to the implants were recorded by using 3D DIC technique. Measurements were repeated after initially increasing torque value to 10 Ncm and then to 15 Ncm, and these steps were repeated for a total of 10 PEEK and 10 Ti SBs on both implants ( n = 20). All SBs were then sterilized 25 times by using an autoclave (STATIM 5000 S G4) according to manufacturer's recommendations and all measurements were repeated. Linear displacements on three axes were calculated for each SB with increasing torque values (from 5 to 10 Ncm and from 10 to 15 Ncm) before and after sterilization. SB displacements within each torque value–sterilization pair were compared by using Mann–Whitney U test, whereas Wilcoxon signed‐rank test was used to compare SB displacements within each material–torque value pair between conditions and within each material–sterilization pair between torque values (α = 0.05).
Results
On x‐axis, PEEK SBs had higher displacements than Ti SBs ( p < 0.001), whereas sterilization ( p ≤ 0.028) and 15 Ncm torque application ( p ≤ 0.006) led to higher displacements of PEEK SBs. On y‐axis, PEEK SBs had higher displacements than Ti SBs with 15 Ncm torque application ( p ≤ 0.033). A total of 15 Ncm torque‐applied PEEK SBs and 10 Ncm torque‐applied Ti SBs had higher displacements after sterilization ( p ≤ 0.028). Application of 15 Ncm torque led to higher displacements regardless of the material ( p ≤ 0.002). On z‐axis, PEEK SBs had higher displacements ( p ≤ 0.015), except for 10 Ncm torque‐applied sterilized SBs ( p = 0.102). With 10 Ncm torque application, sterilization decreased the displacement values of PEEK SBs ( p = 0.044). Greater displacements were observed with 10 Ncm torque‐applied Ti SBs before sterilization and 15 Ncm torque‐applied PEEK SBs after sterilization ( p ≤ 0.033).
Conclusions
Axial displacement of SBs was affected by material type, torque value, and sterilization. Ti SBs mostly had lower displacements than PEEK SBs. Application of 15 Ncm torque to tested PEEK SBs should be refrained from and a calibrated tightening tool may enable the application of 10 Ncm or lower torque values for lower displacements. Sterilization generally increased PEEK SB displacements.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Private Practice, Med Dis Dental Clinic, Ankara, Turkey.
- Department of Prosthodontics, Faculty of Dentistry, Ankara University, Ankara, Turkey.
- Division of Restorative and Prosthetic Dentistry, The Ohio State University College of Dentistry, Columbus, Ohio, USA.
- Department of Reconstructive Dentistry and Gerodontology, School of Dental Medicine, University of Bern, Bern, Switzerland.
- Department of Prosthodontics, Faculty of Dentistry, Istinye University, İstanbul, Turkey.
- Department of Mechanical & Aerospace Engineering, The Ohio State University College of Engineering, Columbus, Ohio, USA.
ORCID-Profile: Emre Diker, Hakan Terzioglu, Diogo N M Gouveia, Mustafa Borga Donmez, Jeremy Seidt, Burak Yilmaz
Zitieren
APA
Diker, E., Terzioglu, H., Gouveia, D. N. M., Donmez, M. B., Seidt, J., & Yilmaz, B. (2023). Effect of material type, torque value, and sterilization on linear displacements of a scan body: An in vitro study. Clinical Implant Dentistry and Related Research. https://doi.org/10.1111/cid.13187
Vancouver
Diker E, Terzioglu H, Gouveia DNM, Donmez MB, Seidt J, Yilmaz B. Effect of material type, torque value, and sterilization on linear displacements of a scan body: An in vitro study. Clinical Implant Dentistry and Related Research. 2023. doi:10.1111/cid.13187
BibTeX
@article{Diker2023,
author = {Diker, Emre and Terzioglu, Hakan and Gouveia, Diogo N M and Donmez, Mustafa Borga and Seidt, Jeremy and Yilmaz, Burak},
title = {Effect of material type, torque value, and sterilization on linear displacements of a scan body: An in vitro study},
journal = {Clinical Implant Dentistry and Related Research},
year = {2023},
doi = {10.1111/cid.13187}
}
EUVIMED