Increased brain coverage and efficiency when measuring current-induced magnetic fields by use of simultaneous multi-slice echo-planar MRI

Teresa Cunha iD, Fróði Gregersen, Lars G Hanson iD, Axel Thielscher iD

PLOS One · 2026

DOI: 10.1371/journal.pone.0341731 ↗

Kurz gefasst

Fachartikel in der Zeitschrift PLOS One, 2026. Verfasst von Teresa Cunha und 3 weiteren Autor:innen. Beteiligte Einrichtung: Section for Magnetic Resonance, DTU Health Tech, Technical University of Denmark, Kgs Lyngby, Denmark. Hauptthemen laut MEDLINE: Brain, Echo-Planar Imaging, Magnetic Fields. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Lundbeck Foundation, Deutsche Forschungsgemeinschaft.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Magnetic resonance current density imaging (MRCDI) can non-invasively validate electric field simulations in volume conductor head models. Weak electric currents are injected using scalp electrodes while measuring the MR phase perturbations caused by the tiny magnetic fields (1–2 nT) induced by the current flow in tissue. MRCDI generally has a low signal-to-noise ratio, making it susceptible to technical imperfections and physiological noise. In this technical note, we tested and optimized simultaneous multi-slice (SMS) EPI for time-efficient and robust brain MRCDI.

Methods

MRCDI data was acquired in a phantom and five human brains using SMS-EPI optimized for measuring current-induced phase perturbations. Multiband factors and interslice gaps were systematically varied and the resulting image quality assessed. In particular, the impact of interslice signal leakage on the measured phase was tested.

Results

Current-free acquisitions showed the expected noise amplification with decreasing interslice distances. However, physiological noise generally dominated the human data, masking potential SMS-related penalties and making the overall noise levels identical to single-slice EPI for interslice gaps of at least 12 mm and multiband factors between 3 and 5. Upon application of electric currents, the phantom data revealed subtle artifacts for multiband factors 5 and 6, even for large gaps. Nevertheless, artifacts were absent in the human brain for multiband factors up to 5, where the performance of SMS-EPI approached that of single-slice measurements for sufficient interslice distances.

Conclusion

Optimized SMS-EPI with multiband factors up to 5 and minimum interslice gaps of 12 mm performs on par with single-slice EPI, making it attractive for increasing brain coverage in MRCDI.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrainEcho-Planar ImagingMagnetic FieldsHumansPhantoms, ImagingSignal-To-Noise RatioImage Processing, Computer-Assisted

Autor:innen und Institutionen

  • Section for Magnetic Resonance, DTU Health Tech, Technical University of Denmark, Kgs Lyngby, Denmark.
  • Danish Research Centre for Magnetic Resonance, Department of Radiology and Nuclear Medicine, Copenhagen University Hospital - Amager and Hvidovre, Copenhagen, Denmark.

ORCID-Profile: Teresa Cunha, Lars G Hanson, Axel Thielscher

Förderung

  • Lundbeck Foundation
  • Deutsche Forschungsgemeinschaft

Zitieren

APA

Cunha, T., Gregersen, F., Hanson, L. G., & Thielscher, A. (2026). Increased brain coverage and efficiency when measuring current-induced magnetic fields by use of simultaneous multi-slice echo-planar MRI. PLOS One. https://doi.org/10.1371/journal.pone.0341731

Vancouver

Cunha T, Gregersen F, Hanson LG, Thielscher A. Increased brain coverage and efficiency when measuring current-induced magnetic fields by use of simultaneous multi-slice echo-planar MRI. PLOS One. 2026. doi:10.1371/journal.pone.0341731

BibTeX
@article{Cunha2026,
  author = {Cunha, Teresa and Gregersen, Fróði and Hanson, Lars G and Thielscher, Axel},
  title = {Increased brain coverage and efficiency when measuring current-induced magnetic fields by use of simultaneous multi-slice echo-planar MRI},
  journal = {PLOS One},
  year = {2026},
  doi = {10.1371/journal.pone.0341731}
}
Kostenlos starten →